SR Instruments SRScales SRV713W Скачать руководство пользователя страница 12

Model SRV713W Series 
Operating and Service Manual 
Part No. MAN713W_210526 

 

Page 12 of 14 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

     

  

Copyright 2021

 

S

Instruments, Inc. 

 

CALIBRATION (Cont’d) 

 

STEP 10:

  When the display reads “

FULL

”, place the weight from Step 1 on the platform and press the 

DISPLAY 

button. 

STEP 11:

  Press the 

DISPLAY

 button once more to save.   To exit without saving, press 

S4 

to select 

QUIT

”, then press the 

DISPLAY 

button

 

STEP 12:

  When finished making adjustments, remove all weight from the scale.   

STEP 13:

 

 

Replace the display box cover and securely tighten screws.  Re-attach display to platform and 

securely tighten screws. 

NOTE: TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS.

 

STEP 14:

  Press and hold the 

DISPLAY 

button to zero the platform.  When the display indicates “

0.0

”, 

place the calibrated weight on the scale again and check against the weight displayed.  If the 
weight  is  not  correct,  recalibrate.  (If  calibration  cannot  be  accomplished,  call  the  Service 
Department.) 

 
 
 
 
 
 
 
 

 

 
 
 

TROUBLESHOOTING 

 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Weight reading is much lower than 
expected.   

Check that the platform is clean underneath, stand on 
each  of  the  four  corners  to  see  if  one  corner  is  not 
weighing correctly.   

For additional information or assistance, telephone our Service Hotline: 1-800-654-6360 

or e-mail:  

[email protected]

 

 
 
 

CAUTION 

The  integrated  circuits  and  semiconductors 
on  the  printed  circuit  boards  may  be 
damaged  by  electrostatic  discharge  (ESD).  
Be sure to use proper handling precautions at 
all times.  

Содержание SRScales SRV713W

Страница 1: ...f 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc S by SR Instruments Inc Ope...

Страница 2: ...URES 2 PACKING CHECKLIST MODEL SRV713W 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 B...

Страница 3: ...Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc PACKING CHECKLIST Model SRV713W Marine Platform Scale DESCRIPTION...

Страница 4: ...check parts against the PACKING CHECKLIST If there are any missing or damaged parts call the Service Hotline at 1 800 654 6360 STEP 2 Place the Platform upside down on the floor and remove the cover...

Страница 5: ...of the subject s weight The SRV713W Marine Platform Scale derives its power from six 6 each AA batteries With the push of a button weight is displayed with a displayed resolution of 0 2 pounds or 0 1...

Страница 6: ...sive cleaners The waterproof seal can also be damaged We recommend mild soap and water for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods lo...

Страница 7: ...seconds AUTO POWER DOWN After five 5 minutes BLUETOOTH ON TIME 5 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA cell batteries CALIBRATION Calibration is trace...

Страница 8: ...tery compartment on the underside of the scale Select the mode for weighing with this Toggle Button Figure 2 LB KG Mode Pushbutton Switch BASIC SYSTEM OPERATION Ensure that the scale is free and clear...

Страница 9: ...replacement is needed an indicator will appear on the display STEP 1 Turn the platform over and remove the four screws on the Battery Cover STEP 2 Remove and replace the six 6 AA cell batteries STEP 3...

Страница 10: ...four strain gauges mounted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electric...

Страница 11: ...weight from Step 1 Use the DISPLAY button to select digit positions and use button S4 to change the value STEP 7 When finished the display will read SAVE Press the DISPLAY button to save or press but...

Страница 12: ...curely tighten screws NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS STEP 14 Press and hold the DISPLAY button to zero the platform When the display indicates 0 0 place the calibrated w...

Страница 13: ...s on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event sh...

Страница 14: ...AN713W_210526 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc S By...