CONFIGURING SEND BUTTON
STEP 1:
Remove the three (3) screws on the left end cover of the display housing. Remove the
cover. Put the scale system into the Calibration Mode by switching the calibration switch on the
display board (Figure 9).
STEP 2:
Press the “
MENU
” key until “
DATA OUT =PRINT
” is displayed
STEP 3:
Press “
UP
” or “
DOWN
” button to select between “
DATA OUT
=NONE
”, “
DATA
OUT
=
” or “
DATA OUT =EHR
”
STEP 4:
Press the “
ENTER
” button to save.
STEP 5:
Switch the scale system out of the Calibration Mode on the display board (Figure 9).
STEP 6:
Place the left end cover on the display housing. Attach using three (3) cover screws.
EHR OUTPUT
Electronic Medical Records (EMR) / Electronic Health Records (EHR) technology is supported by
the i-series scales from SR Instruments. The EMR/ EHR software is not included with scale.
There are many different EMR/ EHR software available and the connectivity of the scale to the
software requires the services of a professional.
A null modem cable is required to communicate with the scale.
Scale output when send button is configured to Electronic Health Records is as follows:
<esc>R<esc>Wnnn.n<esc>Uuu<esc>E
Where
:
R is read
W is weight
nnn.n is the weight in Lb or kg
U is units (lb or kg; in or cm)
uu is LB or KG
E is end of packet.