Squibb-Taylor A120MAN Скачать руководство пользователя страница 5

A120MAN IOM

5 of 7

www.squibbtaylor.com         1.800.345.8105

While this information is presented in good faith and believed to be accurate, Individuals using this literature must exercise their independent judgment in evaluating product selection and determining 

product appropriateness for their particular purpose, system requirements and certifications. The manufacturer reserves the right to change product designs and specifications without notice.

Operation Of The A125-45A, A125-60A, A15-45A, & A15-60A Valves

Maintenance

1. See that the manual release is properly connected, works freely and is not worn.

2. Make sure there are no obstructions or debris to block the valve and lever from closing.

3. Make sure that both cables move freely inside the Break-Away Coupling.

CHECK ACTUATOR PERIODICALLY FOR THE FOLLOWING:

Pull on the Loop End of the Cable. The plunger will retract to allow the lever to move. The closing 

spring causes the valve to close. 

The operator must always be aware of where the remote closure controls are located and know 

how to operate the controls if an emergency requires valve closure. When finished, make a habit  

of closing the internal valve from the remote closure point, thus checking to see that the control 

actually is capable of closing the valve and operating correctly.

To Close Valve

Open

To Open Valve

1.  Always open the internal valve before opening any other valves in the line.

2.  Move the lever to the half-open (rapid equalization) position (Operational Schematic Figure 1, View #2 - A120, 

 

     A125, A15 IOM) to equalize pressure. When the main poppet clicks open, move the operating lever fully open.

3.  Open other line valves slowly to avoid sudden surges which could slug the excess flow valve shut.

4.  If the excess flow valve does close, close the nearest downstream valve. Move the internal valve’s operating 

 

     lever back to the rapid equalizing position (outlined in step 2) and wait for the valve to click open. Then move 

 

     the operating lever fully open and slowly open the downstream valve.

Closed

       Since there is strong spring force on the operating lever, avoid getting in the way of the lever     

 

       as it moves to the closed position. Failure to do so could result in personal injury.

!

WARNING

      DO NOT Force open lever against NH3 Nurse Tank flow before pressure is equalized on each side of the main valve disk. 

      DO NOT Modify the Cable/Carabinier Assembly. Could cause Injury or Death! Only Use Squibb Taylor Replacement Parts.

!

CAUTION

  

ONLY PERSONNEL TRAINED IN THE PROPER PROCEDURES, CODES, STANDARDS & REGULATIONS OF THE ANHYDROUS AMMONIA                          

 

  INDUSTRY SHALL PERFORM

 

MAINTENANCE ON THIS EQUIPMENT. Before starting any type of maintenance, close off the A120MAN 

 

  valve(s) & remove all Anhydrous Ammonia (NH3) pressure from the Outlet of the Internal Shutoff Valve. If maintenance or repairs are to   

 

  be made on the internal valve(s), refer to the Instruction Manual for the particular valve model. An A120MAN manual release that has

  been disassembled must be tested for proper operation before being returned to service. 

!

WARNING

Assembled Valve & Manual Release is now an Emergency Shutoff Valve

for Anhydrous Ammonia (NH3) Nurse Tank.

•  Test For Proper Operation @ A Minimum Of Once A Month & Daily During Field Application Periods.

•  Always Test Operation Before Sending To Field For Use.

Содержание A120MAN

Страница 1: ...LL SQUIBB TAYLOR 800 345 8105 WITH ANY QUESTIONS BEFORE PROCEEDING KEEP THIS DOCUMENT WITH THE PRODUCT UNTIL IT REACHES THE END USER IMPORTANT USER SAFETY RESPONSIBILITY STATEMENT FOR SQUIBB TAYLOR PR...

Страница 2: ...e valve Start threading the screws into the valve 6 Test the Open operation of the Manual Release Move the Lever to the right until it stops release it The plunger on the Manual Release should have re...

Страница 3: ...fore Sending To Field For Use To Open Valve 1 Always open the internal valve before opening any other valves in the line 2 Move the lever to the half open rapid equalization position Operational Schem...

Страница 4: ...tart threading the screws into the valve 6 Test the Open operation of the Manual Release Move the Lever to the right until it stops release it The plunger on the Manual Release should have retracted i...

Страница 5: ...lize pressure When the main poppet clicks open move the operating lever fully open 3 Open other line valves slowly to avoid sudden surges which could slug the excess flow valve shut 4 If the excess fl...

Страница 6: ...the Break Away Insert Manual Release Cable 6 Discard the Break Away Insert Inspect the Crimped Ends of the Cables for wear or deterioration Discard Break Away Insert 8 Replace the Manual Release Cable...

Страница 7: ...111911 A125 60AM Angle 60 GPM 111913 A125 60SM Straight 111916 A15 45AM 1 1 2 MNPT 1 1 2 FNPT Angle 45 GPM 111918 A15 45SM Straight 111921 A15 60AM Angle 60 GPM 111923 A15 60SM Straight WARNING For An...

Отзывы: