9
Maintenance
Sensor
If the tool does not sound an alarm in the presence of a known leak, the sensor
may need to be replaced . Refer to
Replacement Parts
for the sensor part
number .
1 . The sensor is located on the end of the probe tip . Place the power switch in
the OFF / CHARGE position, grip the end of the probe tip and pull the sensor
straight out of the sensor socket .
2 . Align the pins in the new sensor with the receptacles in the sensor socket,
and push the sensor into place .
3 . Test the tool in the presence of a known leak . If it still does not alarm,
immediately discontinue using the tool and return it to TIF for service . Refer to
the
Warranty
section of this manual .
Disposal of Batteries
Batteries are recyclable. Do not incinerate or expose batteries to
open flames . Dispose of batteries according to local, state, and
federal regulations .