Operation

1. 

Press the Power button, 4, to turn 1205CXB on.

2. 

Connect a cable to the BNC connector, 2.

3. 

Press the 

CFG

 button, 5, repeatedly until VOP is highlighted in the 

C

on

F

i

G

 submenu, 24.

4. 

Use the arrow buttons, 11, to set the VOP% to match the VOP% of the 
cable. This can usually be found in the cable’s datasheet under VOP, 
Velocity of Propagation or Dielectric.

5. 

Press the 

CUR

 button, 6, to highlight 

Cursor 1

, 15, and if necessary, use 

the left and right arrow buttons to position 

Cursor 1

 at the start of the 

launch pulse, 17.

6. 

Press the 

CUR

 button again to highlight 

Cursor 2

, 16, and use the left and 

right arrow buttons to position Cursor 2 at the start of the reflected pulse, 
18, as shown in the Display diagram.

7. 

Read the distance, 20, or time, 19, to the discontinuity.

8. 

Additional functions, such as changing measurement units, setting 
date / time, loading a cable from memory, and saving and recalling 
waveforms, can be accessed via the 

MENU

 button, 8. For details, please 

refer to the 1205CXB Operation Manual.

9. 

Press the Power button, 4, to turn 1205CXB off.

Waveforms

The display of the 1205CXB shows a launch pulse at the left hand side of the 
display and a reflected pulse if any cable imperfections are within range.

Move Cursor 2 so that it is positioned at the start of the reflected pulse, 
as shown in the Display diagram. The distance to the imperfection is 
then displayed at the bottom of the display. You can change the units of 
measurement in the 

MENU

.

Open circuit and high impedance series faults will result in a positive (upward) 
reflected pulse. Short circuit and low impedance shunt faults will give a 
negative (downward) reflection.

WaveView

 PC software

Radiodetection’s WaveView PC program is a software package that allows you 
to view, manipulate, print and archive cable signature waveforms produced by 
1205CXB cable fault locator. Visit

 www.radiodetection.com

 for a download 

link and usage information.

Содержание RADIODETECTION 1205CXB

Страница 1: ...1205CXB Time Domain Reflectometer and Cable Analyser User Guide 90 1205CXB UG ENG 02...

Страница 2: ......

Страница 3: ...place after this user manual was published Contact your local Radiodetection dealer or visit www radiodetection com for the latest information about the 1205CXB product family including this guide CA...

Страница 4: ...g documentation the safety of the operator may be compromised Do not expose the equipment to extreme temperatures Store the instrument indoors during extreme hot or cold temperatures and bring the ins...

Страница 5: ...able must contain at least two conductors or one conductor and screen This can be generalized to include other multi conductor set ups such as district heating systems Velocity of Propagation VOP The...

Страница 6: ...Select cursor 1 or 2 7 WAVEform Select move or zoom function for a waveform 8 MENU Select units a pre loaded cable and system settings 9 SAVE Save waveform to 1205CXB s memory or USB 10 LOAD Load a w...

Страница 7: ...able discontinuity 19 Time measurement Time for the pulse to reach the discontinuity 20 Distance measurement Distance along the cable to the discontinuity 21 dBRL measurements dB of Return Loss at cur...

Страница 8: ...cable from memory and saving and recalling waveforms can be accessed via the MENU button 8 For details please refer to the 1205CXB Operation Manual 9 Press the Power button 4 to turn 1205CXB off Wavef...

Страница 9: ...However you can extend your equipment s life by following these care and maintenance guidelines Store the equipment in a clean and dry environment Ensure connection sockets are clean free of debris an...

Страница 10: ...itted modified or used in whole or in part without the prior written consent of Radiodetection Ltd Visit www radiodetection com Follow us on USA Raymond ME Kearneysville WV Canada Vaughan ON Mississau...

Отзывы: