Unlike other functions that user would give direct command on how much to
aspirate in the Source well, Fill Plate function give direct command on how much
liquid to
dispense in the Target well.
Therefore, user only need to set the ll volume, and de ne other features if needed,
the system will automatically calculate how much to aspirate in the Source well.
If the volume is not enough to ll all the de e columns at once, it will go back and
forth from the Source well to the Target well until the lling is complete.
Below shows the Fill Plate menu for 8-Channel tip from 384 Well plate to 96 Well
plate. This function is not allowed using a full head (384-channel, 96-channel or 24-
channel) operation.
The default value for the Maximum Aspirate Volume is depending on the maximum
tip volume. User can only adjust the Maximum Aspirate Volume lower than the
maximum tip volume.
This feature is similar with the Volume Retained in Mix function. But it cannot be edit
as a variable, and the maximum value for this is 25uL regardless on the tip volume.
Содержание Apricot S3
Страница 1: ......
Страница 2: ...2021 SPT Labtech Ltd 2 Notes...
Страница 15: ...2021 SPT Labtech Ltd 15 Step 2 Unscrew silver screws on both sides...
Страница 16: ...2021 SPT Labtech Ltd 16 Step 3 Raise the hood pull out the support rod to place the hood...
Страница 21: ...2021 SPT Labtech Ltd 21 Here s a general overview of how a user will run the apricot S3...
Страница 28: ...2021 SPT Labtech Ltd 28...
Страница 59: ...2021 SPT Labtech Ltd 59...
Страница 71: ...2021 SPT Labtech Ltd 71...