5

How to Add IPA

6

Build Platform Information

MAX

MIN

500 mL

1

2

Minimum

500 mL

Required per tank

Maximum

2 Liters

Required per tank

Isopr

opyl

Alcohol 91%

Recommended

Always wear gloves and 
proper eye protection.

Build Platform Adaptor 

Compatible with Pro 55 and Pro 55s

Open the tank lid and add at least 500mL of IPA 91% or greater 
to each tank. IPA 99% and ethanol may also be used depending 
on preference, availability, and material requirements. 

If the volume of IPA is insufficient to run a cleaning cycle, the 
onboard computer will detect this and notify you.

Keep the tank lid closed to prevent the evaporation and 
contamination of IPA.

When new IPA is added, hydrometers may float 
for a few minutes before sinking

Depending on which SprintRay 3D printer you have, you may 
need to attach a magnet and/or use a fitment adaptor when 
using ProWash S.

How to Install the Magnet

If using Pro95 or Pro55, you’ll need to install the provided 
magnet to enable the platform wash. Thoroughly clean the side 
of the build platform before attaching the magnet sticker.

Using the Build Platform Adaptor

If using a Pro55 or Pro55 S, you’ll need to use the 
included adaptor to enable the platform wash. Place the 
adaptor atop the wash chamber so that the opening is 
on the right and run a cleaning cycle as normal.

1/4 inch

1/2 inch

Center

Magnet

Pro55

 

Print Platform

Place magnet on the right side

Pro95

 

Print Platform

Place magnet on the right side

Label

Center

Magnet

Right Side

Right Side

Printer Model

Pro95
Pro55
Pro95 S
Pro55 S

Install Magnet

Yes
Yes
No, It is embedded
No, It is embedded

Use Adaptor

No
Yes
No 
Yes

Содержание ProWash S

Страница 1: ...aning Brush Section 9 Basket Section 7 Hydrometers Pre installed Section 4 Power Cable USB Cable Section 12 Wash Filter Sections 2 9 Metal Tweezers Section 10 Power Adaptor Magnet Sticker Section 6 Bu...

Страница 2: ...loats will void your warranty IPA Tanks Pour IPA directly into these tanks Each tank has a maximum capacity of 2 liters Tank 1 The IPA in this tank is used for the bulk wash so it will need to be chan...

Страница 3: ...before sinking Depending on which SprintRay 3D printer you have you may need to attach a magnet and or use a fitment adaptor when using ProWash S How to Install the Magnet If using Pro95 or Pro55 you...

Страница 4: ...print batches Dense supports Complex geometries Recommended for Models shorter than 4 Solid models Models for aligners Surgical guides Night guards Do not stack models or crowd them together as this...

Страница 5: ...n the drain Used IPA should be properly disposed of through a hazardous materials collection as required by local laws and regulations WARNING For best results clean the tanks every time you refresh I...

Страница 6: ...mpany The secondary mesh below the removable filter should never be removed WARNING ProWash S has a mesh filter that protects the plumbing from resin and other debris The filter should be cleaned when...

Страница 7: ...and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that inte...

Отзывы: