OPERATING MANUAL

 

MN-SST50 
ED-17-09 

 

 

SPRIANO® TECHNOLOGIES 

- www.spriano.it – [email protected]

 

a TERRANOVA® Srl brand

 - www.terranova-instruments.com - VAT IT07848810151 

Factory & Sales: Via Gramsci 1 – 26827 Terranova dei Passerini (LO) – Italy 
Head Office: Via Rosso Medardo 16 – 20159 Milano – Italy Ph: +39 0377 919119 – Fax: +39 0377 855720 

 

Pa

g.

 

5

/4

8

 

 
1.3

 

D

EFINITIONS 

A

ND 

S

YMBOLS

 

In this manual are used the following symbols and labels 
 

 

Equipment for use in potentially explosive atmospheres 

II 1G 

Equipment  Group  II  Category  1G,  suitable  for  potentially  explosive  Zone  0  area 
(gases, vapours or mists)

 

and with redundancy for zone 1 and 2 

Ex ia 

Equipment protection by intrinsic safety, level of protection "ia" 

IIC 

Group IIC apparatus, fit for substances of group IIC, IIB and IIA 

 

Temperature Class of the transmitter (maximum surface temperature) 

T6, 
T5, 
T4 

 

only when used in the ambient temperature range –40°C to 40°C 

 

only when used in the ambient temperature range –40°C to 55°C 

 

only when used in the ambient temperature range –40°C to 80°C 

Ga 

Equipment with equipment protection level (EPL) “Ga” 

 

 

W

ARNING

Warning

 

label

,  used  to  warn  about  hazardous  materials,  locations,  or  objects, 

including electric currents, poisons, and other things. 
 

 

 

P

ROHIBITION

Prohibition

 

label

, used to indicate something is not permitted. 

 

 

 

M

ANDATORY

Mandatory label

, used to set obligation tasks to be performed by the user. 

 

 

 

I

NFORMATION

Information  label

,  used  to  set  general  information  for  the  user  or  warn  about 

hazards for the equipment. 
 

Содержание SPRIANO Series

Страница 1: ...OPERATING MANUAL MN SST50 ED 17 09 SST50 TORQUE TUBE LEVEL TRANSMITTER ORIGINAL INSTRUCTIONS...

Страница 2: ...on of the harmonised standards EN 60079 0 2012 A11 2013 EN 60079 11 2012 The Notified Body NB 1354 Technicka inspekcia a s performed Module D Annex IV Directive 2014 34 EU and issued the Production Qu...

Страница 3: ...quirements 7 3 3 Mechanical Installation 8 3 4 External Mounting Displacers 8 3 5 Internal Mounting Displacers 10 3 6 Top Part Orientation 11 3 7 Electronic Head Orientation 13 3 8 Electrical Connecti...

Страница 4: ...solutions For more information or for particular problems not considered in this manual please address to our technical office The warranty period is the one contemplated in our general servicing con...

Страница 5: ...protection by intrinsic safety level of protection ia IIC Group IIC apparatus fit for substances of group IIC IIB and IIA Temperature Class of the transmitter maximum surface temperature T6 T5 T4 onl...

Страница 6: ...to display the data on the wide LCD display The transmitter series SST50 measures liquid levels density and interface levels in closed or open vessels It converts buoyant force exerted by a displacer...

Страница 7: ...owing considerations Check whether Transmitter s operating conditions are within the limits as reported in the nameplate technical specifications sheets and or label Make sure that the operating condi...

Страница 8: ...department In case of unstable fluids inform technical department In case of damages inform technical department Inspection frequency consideration should be given to applicable national legislation...

Страница 9: ...59 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 9 48 Strictly follow the bellow sequence Figure 3 remove the adhesive cover from flanges 1 set the Transmitter on plant remove the bottom locki...

Страница 10: ...led coaxially with the displacer and fixed either to vessel or to Transmitter flanged connection The well is made from tubing or pipe with an inside diameter 50 80 large than the diameter of the displ...

Страница 11: ...1 48 3 6 TOP PART ORIENTATION Before disassembly the TOP PART strictly follow the Step 1 Figure 17 Step 2 Figure 18 and Step 4 Figure 19 indicated in TRANSPORT SECURITY OPERATIONS Step 1 Figure 5 once...

Страница 12: ...e Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 12 48 Step 2 Figure 6 rotate the top part in the desired direction then slowly put down it ensuring the coaxiality be...

Страница 13: ...D ORIENTATION In order to rotate the electronic head in the desired position you must follow these passages Figure 7 unscrew the 4 screws A and remove the housing cover B loosen the fixing nut C with...

Страница 14: ...hing WARNING The threaded opening that is not used must be sealed using a suitable cap The caps included in the delivery are only for protection during transport Figure 8 electrical connection INFORMA...

Страница 15: ...ia protection The electronic transmitters totally comply with the HART protocol specification Revision 6 0 so they include remote process variable interrogation parameter setting and diagnostics The d...

Страница 16: ...Head Office Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 16 48 INFORMATION Ref to Figure 10 only one transmitter could be connected to each barrier Ref to Figure 1...

Страница 17: ...assification of the zone and for the presence of flammable substances in the plant see MARKING The safety essential requisite against the risk of explosion in the classified areas are fixed from the E...

Страница 18: ...sociated apparatus has to be made based on the maximum input parameters of the transmitter Maximum input parameters of the equipment relative to the intrinsically safety Parameter Tamb 60 C Tamb 80 C...

Страница 19: ...rthing systems In the case where the plant devoid of earthing it is necessary to connect the earth wire to the terminal block through the cable that carries the signal and the power supply Figure 8 Ca...

Страница 20: ...0 Pag 20 48 Figure 14 earthing in the housing WARNING The hose must not be considered as a protection against electric shock for the protection of security but as protection against electromagnetic in...

Страница 21: ...put current URV it is referred to the level value which gives 20mA output current PV will be the measurement of the level of the fluid on the displacer in of the displacer length or in engineering uni...

Страница 22: ...ode 20 4mA output The direct mode is obtained by setting the lower range point with the lowest value of the range e g 0 mm o 0 kg l and by setting the upper range point with the highest value e g 35 c...

Страница 23: ...display is in numerical entering mode the cursor will blink To move the cursor press the button to raise the value of the digit press the button Press the OK button to confirm use instead the MENU bu...

Страница 24: ...the actual reference input to the zero display measure The transmitter will then show 0 at display after the confirmation by OK 4mA input automatic setting zero SET LRV At button pushing the transmit...

Страница 25: ...nd the firmware revision are shown Password setting PASSWORD Allows to insert a new password for the transmitter To be used with care 0000 9999 Expert command Manual zero elevation ZERO ELEV disabled...

Страница 26: ...the steps indicated in this chapter Failure to comply with this section will permanently damage the instrument Before moving the Transmitter strictly follow the bellow sequence of instructions Step 1...

Страница 27: ...ice Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 27 48 Step 2 Figure 18 screw for seven turns the stop dowel F in order to block the sensor mechanism Step 3 Figure...

Страница 28: ...ni LO Italy Head Office Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 28 48 Step 4 Figure 19 ensure the top part with appropriate equipment in the indicated hanging...

Страница 29: ...AT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 29 48 PROHIBITION Never...

Страница 30: ...as a handle or step Please refer to the following pages scheme for troubleshooting In any case contact our technical department for help Condition Potential source Solution The transmitter doesn t tur...

Страница 31: ...ble to access the menu Wrong forgotten password Contact us for help 6 1 DISPLAY ERROR CODES Type of error Description OVR OVER RANGE this message is shown if PV is greater than URV by 20 of the span U...

Страница 32: ...0377 855720 Pag 32 48 6 2 SPARE PARTS Together with the request of spare parts it is required to indicate Serial Number of the Transmitter to control that the correct components will be delivered Ite...

Страница 33: ...va instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo 16 20159 Milan...

Страница 34: ...8 6 3 DISPLACER SUBSTITUTION Before disassembly the TOP PART strictly follow the Step 1 Figure 17 Step 2 Figure 18 and Step 4 Figure 19 indicated in TRANSPORT SECURITY OPERATIONS Step 1 Figure 23 once...

Страница 35: ...Via Rosso Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 35 48 Step 2 Figure 24 remove the top locking device A C D then unhook the old displacer insert a new displacer into t...

Страница 36: ...ero offset is 4 there are two ways to operate o if the offset has a positive value Ex 6 then you might loosen the nut N with a wrench CH 6 and rotate the screw M from left to right using a wrench CH 7...

Страница 37: ...ts or replace any components inside the electronic enclosure with the power supply turned on To avoid injury to health or may damage the equipment always turn the power supply to off and ground the ci...

Страница 38: ...so Medardo 16 20159 Milano Italy Ph 39 0377 919119 Fax 39 0377 855720 Pag 38 48 Step 2 Figure 27 maintaining in position the washer and the nut C insert the new electronic head fix the nut C using key...

Страница 39: ...nscriptions showing the characteristic data of the instrument Refer to Figure 28 the Transmitter is provided with the following information 1 Name and address of the manufacturer 2 CE marking followed...

Страница 40: ...nominal range in factory Low upper range values they can be set within nominal range providing that the span minimum span Damping digitally adjustable from 0 to 60 sec minimum response time 0 1 sec R...

Страница 41: ...ard 2 terminals for signal wiring up to 1 5mm 14 AWG Connection for ground and cable shield Mounting see DIMENSIONAL DRAWINGS Net weight see REFERENCE TABLE DESIGNED CONSTRUCTED AND TESTED TO ASME B31...

Страница 42: ...955 1158 1564 1869 2174 2479 2763 3393 Weight Kg 33 35 37 41 44 47 50 53 59 TL A 601 855 1058 1464 1769 2074 2379 2683 3293 B 543 797 1000 1406 1711 2016 2321 2625 3235 C 178 305 407 610 762 915 1067...

Страница 43: ...va instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo 16 20159 Milan...

Страница 44: ...ova instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo 16 20159 Mila...

Страница 45: ...priano terranova instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo...

Страница 46: ...spriano terranova instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medard...

Страница 47: ...spriano terranova instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medard...

Страница 48: ...anova instruments com a TERRANOVA Srl brand www terranova instruments com VAT IT07848810151 Factory Sales Via Gramsci 1 26827 Terranova dei Passerini LO Italy Head Office Via Rosso Medardo 16 20159 Mi...

Отзывы: