Chapter 4: MicroLead Placement & System Set-Up
42
| SPRINT
®
PNS System – Clinician Instruction for Use [L0090-MAN-000 Rev F]
4.4.2) Establishing Therapy Settings using the Clinical Programmer Tablet
If the Clinical Programmer Tablet is not connected to the Pulse Generator, turn on the tablet and
connect to the Pulse Generator now (see Section 4..2.2 for instructions on connecting to the Pulse
Generator).
If the Clinical Programmer Tablet was just used to test and confirm lead placement(s) and the Set
Min/Max button has been pressed, the first step below (selected View/Edit Program) can be
skipped.
Select View/Edit Program to begin setting stimulation minimum
and maximum levels.
Single Lead
Select Start Stim to begin setting minimum and maximum
stimulation levels.
To set the minimum stimulation the patient is allowed to reach,
use the
and
buttons to decrease or increase stimulation
until the desired minimum level is reached. Then tap the number
below the word MIN to set the minimum at the currently
displayed stimulation intensity level.
Immediately after placing the leads, the maximum stimulation
level is set automatically by the system so that patients will have
access to higher stimulation levels that may be useful during
therapy. See Note below on setting the maximum intensity
manually.