10
Meaning of lights
OC (green)
ICL (orange)
Meaning
Continuously
illuminated
-
Standby
Flashes rapidly
-
Measuring mode
Flashes slowly
-
Bootloader runs (exception)
Continuously
illuminated
Flashes slowly
Battery charging
Continuously
illuminated
Continuously
illuminated
Battery is fully charged
Flashes rapidly
Flashes rapidly
Battery error
Continuously
illuminated
Flashes twice
With warning sound: Shutdown, battery
empty
No reaction
No reaction
Imminent deep discharge, protection
mode
Regular use
If the device is switched on and can be used without malfunctions, it is in state
or
.
Charging
In the states
,
(charging) the measuring device is not ready to receive any commands over
the buttons or the software.
If possible, quit the charging process only in state
.
Error conditions
If the state
occurs, contact the manufacturer or distributor for customer
support.
If the state
occurs, the device should be charged as soon as possible to
prevent deep discharge and subsequent damage of the battery.
If the measuring device can no longer be switched on, deep discharge is
imminent and the device is in protection mode (
). If possible, immediately
charge the device. Despite a connection to a power source the protection mode
may persist for some time until the device returns to the charging state
.
Prior to the first use
First fully charge the battery of the device with the charging cable connected to
an USB port or power outlet.
Switching on and off
Push [ENTER], i.e. the middle button, for 3 seconds to switch the device on and off.