Input Signal Sensitivity Adjustment (r1)

The R1 Input Signal Sensitivity Adjustment is a valuable tool that, in addition 

to SP714-S2i being directly mounted to the turbine meter, allows effective 

operation in noisy environments. The R1 is a potentiometer that sets the 

amplitude of the input signal (from the pickup coil) that the SP714-S2i will 

accept as a valid pulse. 

Adjusting R1 allows the SP714-S2i to differentiate between the input signal 

and any noise in the system. Turning the R1 clockwise decreases the 

amplitude of the input signal required for a valid pulse, and turning the R1 

counterclockwise increases the required amplitude.

The detection level of the R1 potentiometer is set at approximately 17 mV 

rms at the factory. A red varnish is applied to the potentiometer screw to seal 

the factory setting.

settInGs & dIaGnostIcs

SW1

S1

J U 4

D1

D2

GND

TB2

S2

R1

TB1

ON

ON

1 2

GND

  1

R1

Input Sensitivity

Adjustment

sW2

Diagnostic Test 

Oscillator

D1 & D2

Diagnostic LEDs

Diagnostic Operational LeDs (D1 & D2)

There are two Diagnostic Operational LEDs (D1 and D2) on the SP714-S2i PC board that indicate the current 

operational status of the preamplifier. The green LED (D1) indicates the power supply status. When the green LED is 

on, the power supply is sufficient. When the LED is dim or flashing, the preamplifier does not have sufficient power.
The red LED (D2) indicates the status of the signal traveling through the preamplifier. The red LED flashes at the 

same rate as the frequency as the signal passing through the SP714-S2i. If the LED is not flashing, no signal is 

passing through the amplifier. If the signal frequency is above 40 Hz, the light will flash very quickly and the LED will 

appear to be constantly illuminated. If the input into the SP714-S2i stops on a positive portion of the signal pulse, the 

LED will be illuminated constantly.

Diagnostic Test Oscillator (SW2)

The SW2 Diagnostic Test Oscillator verifies that the SP714-

S2i is operational. To activate the test, press the SW2 push 

button. A 10 Hz signal will be introduced into the SP714-

S2i signal output pins (TB1). For best results, use the SW2 

Diagnostic Test Oscillator without the presence of the pickup 

coil signal.

If an input signal from the pickup coil is present, the 10 Hz 

signal from the SW2 can disrupt the SP714-S2i output signal 

and an error will be detected by the pulse acquisition device.

Do Not Use the SW2 During Operation

The D1 and D2 LEDs are inoperable in the 2-wire Standard configuration.

Polarity of the connector at the 

pickup coil is not important.

maintenance - cPu board replacement

The CPU board (PN 85025) is the only SP714-S2i part available for purchase. If you are replacing the 85025 CPU 

board, the wiring harness, which connects the pickup coil and to the SP714-S2i CPU board, must be unplugged from 

the old CPU board and plugged into the replacement board, then wired appropriately.

To replace the 85025 CPu board:

1.  Remove the SP714-S2i housing cover.
2.  Pull the CPU board from the housing, disconnect the 

wires to the power supply and the pulse acquisition 

device, and unplug the wiring harness.

3.  Plug the wiring harness into the replacement CPU board, 

wire the replacement CPU board to the power supply 

and the pulse acquisition device, set the board into the 

housing, and replace the housing cover. 

See pages 7-10 for 

detailed instructions.

Wiring Harness

SW1

S1

J U 4

D1

D2

GND

TB2

S2

R1

TB1

ON

ON

1 2

GND

  1

Input signal 

connector

11

Содержание SP714-S2i

Страница 1: ...Liquid Controls Group An IDEX Fluid Metering Business Sponsler SP714 S2i Pulse Amplifier Installation Parts Operation EM700 10...

Страница 2: ...strictly in the hands of the owner operators of the equipment Neglect of that responsibility is not within the control of the manufacturer NOTICE The most current English versions of all Liquid Contr...

Страница 3: ...fire department state and local codes to ensure adequate preparation Read this manual as well as all the literature provided in your owner s packet Save these instructions for future reference Failur...

Страница 4: ...40 to 176 F 40 to 80 C Storage 76 to 257 F 60 to 125 C Specifications Input Voltage 6 28 VDC 83 mA maximum 24 VDC Protected against polarity reversal Signal Input Frequency 2 kHz maximum over operatin...

Страница 5: ...y voltage Vcc 4 28 VDC Total System Resistance 500 minimum 28 VDC Output Voltage Logic HIGH Vcc 1 VDC Logic LOW 0 25 VDC Output Current Logic HIGH 8 to 60 mA DC Logic LOW 8 mA DC DC power supply volta...

Страница 6: ...eratures above 70 C use field wiring suitable for 20 C above maximum ambient temperature II Suitable for use in surface not mine installations 2 G High level of protection is provided against flammabl...

Страница 7: ...uration translates the output signal from the turbine flowmeter s pickup coil into a current value 2 wire connections are effective transmitting the signal over long distances The 3 wire TTL Sourcing...

Страница 8: ...llator SW2 and or check the diagnostic operational LEDs D1 D2 See page 11 9 Coil the wires underneath the CPU board and place the board inside the housing ensuring the wires are not pinched and replac...

Страница 9: ...using the information below 7 Wire the SP714 S2i to the pulse acquisition device and DC power supply as indicated in the schematic 8 If necessary adjust the input signal sensitivity R1 activate the di...

Страница 10: ...n the schematic 8 If necessary adjust the input signal sensitivity R1 activate the diagnostic test oscsillator SW2 and or check the diagnostic operational LEDs D1 D2 See page 11 9 Coil the wires under...

Страница 11: ...equency is above 40 Hz the light will flash very quickly and the LED will appear to be constantly illuminated If the input into the SP714 S2i stops on a positive portion of the signal pulse the LED wi...

Страница 12: ...rive Lake Bluff IL 60044 847 295 1050 SAMPI Via Amerigo Vespucci 1 55011 Altopascio Lucca Italy 39 0583 24751 IDEX Fluid and Metering Pvt Ltd Survey No 256 Alindra Savli GIDC Manjusar Dist Vadodara 39...

Отзывы: