15
A Main display
B
Spectrum
settings
C
Spectrum resolution settings
The spectrum resolution equals the maximum frequency
(1000 Hz) divided by the number of lines of resolution. A
higher number of lines allows more detailed spectrums;
the drawback is slightly longer measurement times.
By activating the ‘AVERAGE’ parameter, VibChecker will
perform four measurements, and then display an average
of the measurement results.
To set a number of spectrum lines:
1. Go to the
Main
display (A).
2. Use the arrow keys to navigate to the
Spectrum set-
tings
icon (B), then press
ENTER
.
3. In the
Spectrum settings
display, use the
UP
/
DOWN
arrow keys to highlight the checkbox for your preferred
number of lines and press
ENTER
to mark it.
4. If averaging should be used, mark the ‘AVERAGE’
checkbox.
5. To save the settings, use the
DOWN
arrow key to
highlight the
SAVE
checkbox and press
ENTER
.