![Spirit Healthcare TEE2+ Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/spirit-healthcare/tee2/tee2_owners-booklet_1348585056.webp)
56
www.i-sens.com
Within ±0.83 mmol/L and Within ±15 %
599/600 (99.8 %)
System accuracy results for glucose concentrations between
2.2 mmol/L and 26.0 mmol/L
Precision:
The precision studies were performed in a laboratory
using TEE2 Plus Blood Glucose Monitoring Systems.
Within Run Precision
Blood average
2.3 mmol/L
SD = 0.1 mmol/L
4.7 mmol/L
SD = 0.2 mmol/L
6.9 mmol/L
CV = 3.4 %
11.4 mmol/L
CV = 3.3 %
17.8 mmol/L
CV = 3.5 %
Between Run Precision
Control average
2.2 mmol/L
SD = 0.2 mmol/L
6.7 mmol/L
CV = 2.5 %
18.8 mmol/L
CV = 4.0 %
This study shows that there could be variation of up to 4.0 %.
Содержание TEE2+
Страница 1: ...Large LCD No Coding Intuitive Meal icons Blood Glucose Monitoring System Owner s Booklet...
Страница 4: ......
Страница 60: ...60 www i sens com MEMO...
Страница 61: ...www spirit healthcare co uk 61 MEMO...
Страница 62: ...62 www i sens com MEMO...
Страница 63: ...www spirit healthcare co uk 63 MEMO...
Страница 64: ...64 www i sens com MEMO...
Страница 65: ...www spirit healthcare co uk 65 MEMO...
Страница 66: ...66 www i sens com MEMO...