Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range results may occur in following situations:
Situations
Do This
•
When the control solution bottle was
not shaken well,
•
When the meter, test strip, or the
control solution were exposed to high
or low temperatures,
•
When the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
•
When the meter is not functioning
properly.
Repeat the
control solution
test by referring
to the 'Notes' on
page 26.
•
When the control solution is past the
expiration date printed on the bottle,
•
When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
•
When the control solution is
contaminated.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
If results continue to fall outside the range printed on the test
strip vial, the TEE2 Test Strip and TEE2 Plus meter may not be
working properly. Do not use your system and contact Spirit
Healthcare 24/7 Freephone Support Line at 0800 881 5423.
Содержание TEE2+
Страница 1: ...Large LCD No Coding Intuitive Meal icons Blood Glucose Monitoring System Owner s Booklet...
Страница 4: ......
Страница 60: ...60 www i sens com MEMO...
Страница 61: ...www spirit healthcare co uk 61 MEMO...
Страница 62: ...62 www i sens com MEMO...
Страница 63: ...www spirit healthcare co uk 63 MEMO...
Страница 64: ...64 www i sens com MEMO...
Страница 65: ...www spirit healthcare co uk 65 MEMO...
Страница 66: ...66 www i sens com MEMO...