background image

RF Power Amplifier PA15W

I. Precautions

Working with RF power amplifiers can be dangerous and even fatal if not handled properly.

Output voltages can reach values greater than 90 V peak-to-peak and can be fatal.  Follow these

steps to avoid damaging the amplifier or inflicting serious injuries.

Connecting & Disconnecting Power Amplifiers

When   connecting   the   power   amplifier,   follow   these   steps  in   order  to   avoid   damaging   the

amplifier or inflicting serious injuries. 

1) Apply the load to the amplifier (make sure a load is ALWAYS present when working with 

power amplifiers).

2) Apply the DC power to the amplifier.

3) Apply the RF input to the amplifier.

      Repeat the steps in reverse order when disconnecting the amplifier.

Power Considerations

Make sure the following considerations have been made before applying power to the amplifier. 

1) Be sure your load can appropriately dissipate the maximum power being applied by the 

amplifier. 

2) When applying an RF signal, work with low duty cycles to limit the power being 

dissipated.  The duty cycle ratio should be below 1% for safe operation. If you would like to 

operate at high duty cycles or in continuous operation, please inquire with SpinCore 

Technologies. 

3) Do not input a sinusoidal RF signal to the amplifier with an amplitude greater than 1 Volt 

peak-to-peak, as values beyond this can damage the amplifier.

http://www.spincore.com

 4 

         2020-02-20

Содержание PA15W

Страница 1: ...RF Power Amplifier PA15W Owner s Manual SpinCore Technologies Inc http www spincore com...

Страница 2: ...chnologies Inc reserves the right to make changes to the product s or information herein without notice RadioProcessor PulseBlaster SpinCore and the SpinCore Technologies Inc logos are trademarks of S...

Страница 3: ...Electrical Specifications 6 3 Typical Performance Data 6 4 PA15W RF Module with Deblanking Circuitry 8 5 PA15W RF Module Enclosure Options 11 1 PA15W RF module enclosed in a single bay enclosure 11 2...

Страница 4: ...o the amplifier 3 Apply the RF input to the amplifier Repeat the steps in reverse order when disconnecting the amplifier Power Considerations Make sure the following considerations have been made befo...

Страница 5: ...conserve power keeps the amplifier cool under typical operating conditions and provides noise reduction during the reception of NMR signals To deblank the power amplifier a TTL pulse needs to be appl...

Страница 6: ...and with the output terminated with 50 Ohms the PA15W provides the performance values contained in Table 2 Table 2 Typical PA15W performance from 1 MHz to 100 MHz Frequency MHz Output Voltage V pk pk...

Страница 7: ...15W RF Power Amplifier 1 5 10 20 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 Output Power vs Frequency Pout W Frequency in MHz Pout W Figure 2 Gain vs Frequency of the PA15W RF Power Amplifier 1 5...

Страница 8: ...V DC Power Input Left screw on terminal block is DC Power which can be identified by the small 1 marked along side it on the board Must be stabilized 12 V DC If voltage applied is higher or lower tha...

Страница 9: ...1 s or so triggering the oscilloscope on the falling edge of the deblanking pulse will help with identifying and capturing the RF pulse on the scope http www spincore com 9 2020 02 20 Figure 5 Represe...

Страница 10: ...re 6 below http www spincore com 10 2020 02 20 Figure 6 Typical performance of the PA15W at the end of pulse Channel 2 is the deblanking signal and channel 1 is the RF output The PA15W was driven by a...

Страница 11: ...rchase For typical enclosures the power input is 90 264 V AC the RF input and output signals are provided on external BNC connectors and the TTL inputs are routed through female DB 9 input connector s...

Страница 12: ...nal specifications include Metal body with streamlined plastic front panel Built in 8x8 cm quiet cooling fan High quality 80 Watt AC input switching power supply Dimensions are approximately 11 x 7 x...

Страница 13: ...io frequency RF digital data acquisition and excitation system For more information please visit http www spincore com products RadioProcessor 4 If you require a custom power level please inquire with...

Отзывы: