Operating Manual
19
1300-87075-1A
Step 3.
Insert Sieve and Container
Retain
Container
Grasp Here
Grasp Here
Sieve
Jar
Step 4.
Slide on Sieve Mechanism
Step 5.
Grasp Firmly and Flip
Warning: Finely pulverized sample always presents a respiratory hazard; a dust mask or other protection
should be worn when working with it.
Grinding jars should be cleaned after use. Methods include wiping with a damp paper towel, blowing out with an
air hose, rinsing with water or alcohol, etc. Most of these methods are not very effective, and blowing the dust
out of a jar with an air hose certainly causes a respiratory hazard. We recommend a 3-minute period of grinding
with a mixture of clean quartz stone, hot water, and soap or laboratory detergent. The residue is easily and
safely washed out of the jar, which then may be dried.
Technicians who are particularly concerned about cross-contamination may adopt the “grind and discard”
technique. After sample A has been ground and the grinding jar nominally cleaned, a portion of sample B is
ground and discarded, and the jar cleaned again. Now it should be possible to grind a portion of sample B
without contaminating it with sample A.
If a cleaned grinding jar will not be used immediately, store in a plastic bag. Hardened steel grinding jars should
always be dried and stored carefully, as they are prone to surface rusting.
9.5 Safety Recommendations
The Planetary Mill is intended for use only by qualified and trained personnel. For questions about the operation,
maintenance or service of the 8200 Planetary Mill, contact us at +1.855.GET.SPEX or +1.732.623.0465.
Every effort has been made to ensure that the Planetary Mill is safe to operate. However, the safety protection
provided by the unit may be impaired if the Planetary Mill is operated in a manner other than what is
described in this manual. In addition, the Planetary Mill should only be used with accessories provided by, or
recommended by, Spex and must be used in the intended manner. Use of accessories not recommended by
Spex may negatively affect the safety protection provided by the unit and may void the warranty.
Do not use the 8200 Planetary Mill with hazardous materials for which the unit was not designed. Be aware
of the hazards of the materials that are being used, particularly in the event of a spill. For instance, use of a
flammable liquid could create a fire hazard if a sample is spilled.