1.0 INTRODUCTION 
 

SPEX SamplePrep Freezer/Mills are laboratory mills that cool unconventional materials to 
cryogenic temperatures and pulverize them to a powder form without thermal degradation.   
Our proven technology uses a dual electromagnetic, grinding chamber that rapidly drives a 
steel impactor back and forth against the two end plugs of the sample vial.  Since the vial is 
securely closed the integrity of its contents is maintained, hazardous or critical samples are 
easily controlled, and cross-sample contamination is eliminated.  The sample chemical 
composition is preserved as a result of the vials being immersed in liquid nitrogen at cryogenic 
temperatures throughout the grinding process.  The 6875D Freezer/Mill expands on this 
technology utilizing dual grinding chambers and two additional precooling chambers for 
increased sample throughput.  The 6875D Freezer/Mill has been re-engineered with a liquid 
nitrogen Auto Fill feature which comes standard in every unit.  The preference to Manual Fill 
the unit with liquid nitrogen is optional.  These unique aspects have made SPEX SamplePrep 
Freezer/Mills the most effective cryogenic mills in the world.  Th

ey are the “mills of first choice” 

for many abnormal materials, or samples whose composition or structure cannot be ground 
using conventional grinding methods. 
 

NOTE:  Do not operate the 6875D Freezer/Mill until you have read the Operating Manual and 

are familiar with the controls and operation.  

 

OPERATING CONDITIONS:

  Do not set up the 6875D Freezer/Mill in an insulated or confined 

space.  The Freezer/Mill should be run on an open countertop, in ambient air, with 
recommended maximum relative room humidity 70% from 5°C - 31°C (40°F - 88°F).  If the entire 
mill is chilled during operation, components such as the gas springs, controller, and display 
screen can fail.  This damage is not covered by the Freezer/Mill warranty.  During operation, 
water vapor in the air will condense on parts of the unit.  For this reason, it is important not to 
place any moisture-sensitive equipment near the Freezer/Mill.

 

 

 

VERY IMPORTANT:

  Liquid nitrogen(LN) not only makes samples brittle through severe chilling 

and 

making them “grindable,” but 

it also cools the magnetic coils which powers the 6875D 

Freezer/Mill.  If the unit is operated in the optional Manual Fill mode the liquid nitrogen level 
should be visually checked during extended runs as well as topped off before every run.   
Operating the unit without liquid nitrogen for a period of one minute will cause the coils to 
become very hot and may sustain permanent damage.  The 6875D Freezer/Mill has a LN sensor 
that will shut down the unit when the liquid nitrogen gets too low to cool the coils.  The 
Freezer/Mill warranty does not cover damage to the coils caused by operating the mill with 
little or no liquid nitrogen.  We recommend operating the unit in Auto Fill mode with LN Tank.  

 

SAFETY

: Liquid nitrogen (LN) can be hazardous.  Its boiling point is -195.8°C (-320.4°F).  When 

working with liquid nitrogen directly or indirectly, the LN Tank valve or hose, or chilled 
Freezer/Mill components, cryogenic gloves must be worn to protect hands.  A face Shield is also 
recommended to protect eyes from possible splashing.  Wear the proper protection equipment 
PPE). 

 

 

Содержание 6875D

Страница 1: ...V 1 65 Liberty St Metuchen NJ 08840 1 855 GET SPEX 438 7739 or 732 623 0465 www spexsampleprep com 6875D LARGE FREEZER MILL OPERATING MANUAL For 6875D 115 and 6875D 230 Freezer Mills this manual is Pa...

Страница 2: ...the work may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording except as may be expressly permitted by the 1976 Copyright Act or in...

Страница 3: ...nel 13 5 3 Changing Settings 17 5 4 Saved Protocols 18 6 0 Operation 20 6 1 Standard Auto Fill 20 6 2 Manually Adding Liquid Nitrogen 21 6 3 Loading Vials into the Freezer Mill 24 6 4 Removing Vials f...

Страница 4: ...o not set up the 6875D Freezer Mill in an insulated or confined space The Freezer Mill should be run on an open countertop in ambient air with recommended maximum relative room humidity 70 from 5 C 31...

Страница 5: ...nce of any vial sample combination depends on sample properties cooling time grinding time desired outcome etc Effective capacity for a given sample is determined by experiment Overfilling a vial grea...

Страница 6: ...ories We recommend storing the packaging materials in the event there is a need to return the unit for warranty service or repairs Grinding vials must be purchased separately and are available in four...

Страница 7: ...RJ45 Ethernet cable connector At the back of the controller is an USB Port to download run data or to upgrade software Caution 4 1 Electrical Connection Plug the electrical cord into the AC Inlet on...

Страница 8: ...scribed and illustrated in Section 5 2 The programmable parameters include Cycles number of grinding periods Precool Time initial chilling of sample Run Time grinding period Cool Time time between gri...

Страница 9: ...e a male 3 8 inch 9 5 mm NPT outlet SPEX transfer hoses come with a matching female 3 8 inch NPT fitting Excessively long transfer hoses are not recommended LN Tanks come in two types high pressure an...

Страница 10: ...Chamber 1 and Chamber 2 of the assembly The Gate keeps the vial s in place during grinding for Chambers 1 and 2 only The vials in the precooling chambers will remain secure and unaffected by the magne...

Страница 11: ...from being interrupted by low liquid nitrogen levels use the Auto Fill mode with LN Tank connected In the optional Manual Fill mode the unit will not run if the proper LN level is not maintained The...

Страница 12: ...the screen with a fingertip or stylus Do not use a sharp point as this can damage or deface the screen 5 1 Logo Screen When the power is switched ON at the back of the mill Please Wait appears during...

Страница 13: ...Chamber 1 will run as indicated by the MONO button color highlighted blue 1 Precool the coil is inactive while the vial cools in liquid nitrogen for 1 minute Samples must be precooled before grinding...

Страница 14: ...or paused in the middle of a grinding program from the button selections to the right of the screen The Auto Fill or Manual Fill buttons are also displayed on Control Panel When a button is selected...

Страница 15: ...tub is filled with liquid nitrogen to the correct level and the lid is securely latched the STATUS BAR will turn green and begin counting down The PRECOOL sequence will start indicating the beginning...

Страница 16: ...indicating by the yellow symbol At the end of each Cool Time period a green check mark is placed next to that field The Rate is 10 cps cycles per second The STATUS BAR top of screen counts down the t...

Страница 17: ...d after the program has stopped then the entire program will start over To pause during a grinding program touch the button The RUN will hold the settings at that moment elapsed time cycle etc To resu...

Страница 18: ...s the last programed run setting that was not saved The Default protocol can be changed without saving the new settings as described in section 5 4 5 4 1 Storing a new program To store the new program...

Страница 19: ...e anything on the screen but returns the display to the SAVED PROTOCOLS screen To run the newly saved protocol touch the protocol to highlight BLUE the name Then touch the LOAD button to send protocol...

Страница 20: ...the Freezer Mill and pressure is relieved through this vent 6 1 Standard Auto Fill When the 6875D Freezer Mill is set up for the first run of the day with a loaded vial s in the grinding Chamber s ma...

Страница 21: ...only be added to 6875D Freezer Mill with the lid open To open the lid hold the lid down with one hand and with the other hand pull the upper end of the latch toward you Then pull the lower end of the...

Страница 22: ...vial and fill the vial 1 3 to 1 2 the volume with sample pieces Typical sample sizes for the various size of vials are shown below in Table 1 Table 1 Typical Sample Sizes Sample Large Vial 6801 6803...

Страница 23: ...aped piece of bone or a tough polymer can act as a wedge to push a moving impactor through the side of the polycarbonate tube 6 2 3 Loading a Sample into a Vial at Room Temperature Attach the end plug...

Страница 24: ...e lid must be closed and latched for the Freezer Mill to run Push the lid down hold it down while engaging the bottom part of the latch and then lock the latch by pushing its top end toward the cabine...

Страница 25: ...st For this reason handle vials with care It is best to open chilled vials immediately after removing them from the Freezer Mill 6 5 2 Emptying a Vial As soon as an end plug is removed from a cold via...

Страница 26: ...samples uses but as soon as they begin to crack they should be discarded Liquid nitrogen can enter through cracks and vaporize when vial is removed from Freezer Mill As the vial warms vaporized LN wi...

Страница 27: ...ources button On the Resources screen as seen in Figure 15 are additional buttons to recall the run history technical support accessories e g vials holders etc and training information e g how to vide...

Страница 28: ...ractors as shown in Figure 17 To open any of the categories listed touch the button to the right of the screen For example pressing the button for the Extractors displays the screen for items availabl...

Страница 29: ...g button on the Resources screen In the Training screen the operating manual instructions and how to use videos are listed To open any of the categories listed on the screen touch the button name as d...

Страница 30: ...en 7 5 Technical Support To access the Technical Support screen touch the Tech Support button on the Resources screen SPEX contact information for the USA Headquarters and Europe Sales Office is liste...

Страница 31: ...ny samples powder materials or liquids are spilled inside the unit during a sample run wait until Freezer Mill is finished for the day the liquid nitrogen has evaporated and the mill has warmed up to...

Страница 32: ...mpletely Close the lid and latch Push the lid closed and latch into place Screen displays LN LEVEL LOW Liquid nitrogen sensor detects inadequate liquid nitrogen level In Auto Fill mode check LN Tank a...

Страница 33: ...and start again Flanged end plug doesn t come out the cold vial End plug tube joint too tight when very cold Allow vial to warm up Blunt end plug doesn t come out room temperature vial Hard to grip W...

Страница 34: ...negligence or misuse 10 1 Product Specifications Every effort has been made to provide complete and accurate product operating information in this Manual However since specifications are subject to ch...

Страница 35: ...ent with the disposal symbol must not be disposed of with general waste See Figure 22 Throughout the European Community guidelines regarding disposal regulations may vary from territory to territory P...

Страница 36: ...e www spexsampleprep com E mail sampleprep spex com In Europe please contact our European Sales Office at SPEX Europe 2 Dalston Gardens Stanmore Middlesex HA7 1BQ United Kingdom Tel 44 0 208 204 6656...

Отзывы: