6. OTHER FEATURES

6.1. Clamp jaw light / Backlit LCD screen

1. Rotate the dial to any position other than off

2. Press and hold the HOLD button for 2 seconds or more

3. The backlit LCD screen will illuminate

4. If the meter is on either of the current settings the clamp jaw light will also illuminate

5.  The backlit LCD screen and clamp jaw light will remain lit for 15 seconds,  

or press the HOLD button for 2 seconds to turn them off

6.2. Auto shut off / Auto shut of disable

1. The meter has an auto shut off feature to preserve battery life

2. The meter will shut off after 15 minutes of non-use

3. Data hold information will not be retained if the auto off feature is enacted 

4.  If the meter is turned off by this feature the operator should rotate the dial back to the  

"OFF" position to resume normal operation

5. The auto shut off feature can be disable by holding down the  "SEL" key when rotating the dial from the off position

6.3. Low battery indicator

1. The meter has a low battery indicator that will alert the user when the voltage is too low to provide accurate readings

2. Replace the batteries when this indicator is displayed (See Section 7, Battery Replacement)

6.4. Data hold

1. The meter has a data hold feature that will retain the measured value on the LCD screen.

2. Press the 

HOLD button momentarily to retain the measured value

3. The      symbol will be displayed while the value is being retained

4. Press the 

HOLD button again to resume normal operation

6.5. Polarity indication

1. The meter displays "-" for negative values, positive is implied

6.6. Overload indication

1. When the input exceeds the measuring range "OL" or "-OL" is displayed on the LCD screen

6.  Press the trigger to open the clamp jaws and place the jaws around a single conductor.  

(Easily accomplished with the use of a line splitter)

7. Turn on the device being tested 

8. The inrush current during the 100ms after start-up will be displayed on the LCD screen 

9. Press the INR/REL button again to clear the display
10. Press and hold the INR/REL button to exit the function

5.9. Relative measurement

1. While taking a measurement press the 

INR/REL key

2. This puts the meter in the relative value measurement mode. 

3. Pressing the 

INR/REL key makes the currently displayed value a reference value in memory. 

4. Subsequent measurements will be displayed as the measured value minus the stored value.

5. The meter does not auto range during this function

6. The meter will display “OL” if the measurement exceeds the range

5-10. Peak Hold

1. The Peak Hold function can be used in either the voltage or current settings

2. The meter will capture and hold the highest and lowest values measured

3. The meter does not auto range during this function

4. Follow the first few steps for the measurement type being performed

5. Press the Peak/Hold button after the meter is set to take the measurement

6. "PMAX" will be displayed on the LCD screen

7. Begin taking the measurement

7. The highest value measured will be displayed on the LCD screen

8. If "OL" is displayed the range will need to be changed manually

9.  To change the range press and hold the Hz% button for two or more seconds  

followed by pressing the Peak/Hold button

10. "PMAX" will be displayed on the LCD screen

11. Retake the measurement

12. If "OL" is still displayed repeat step 9

13.  Once the range is set properly the Peak/Hold button can be used to toggle  

through PMAX, P MIN and PMAX-MIN

I

 peak

T

Содержание DSA600TRMS

Страница 1: ...erminal Input Terminal Peak Button PEAK HOLD HOLD Backlight Clamp Light Dial Frequency Duty Cycle Hz Inrush Current Relative Measurement INR REL 1 Figure 2 Data Hold AC Alternating Current Low Battery...

Страница 2: ...tion manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition Read through and understand the instruction...

Страница 3: ...horoughly before use Measurement categories Over voltage categories To ensure safe operation of measuring instruments IEC61010 establishes safety standards for various electrical environments specifie...

Страница 4: ...0A 0 01A 5 60 600A 1A Input impedance 10M Maximum input voltage 600V AC RMS Frequency range 40 1000Hz AC VOLTAGE RANGE RESOLUTION ACCURACY 6V 0 001V 1 2 5 60V 0 01V 0 8 5 600V 0 1V Duty ratio Clamp Ja...

Страница 5: ...ity Do not use test leads in applications that exceed their CAT rating Check the CAT rating of the test lead and refer to Figure 3 4 5 Checking the switch setting operation Confirm the dial is set to...

Страница 6: ...nstall the test leads 2 Rotate the dial to the position 3 Confirm OL is displayed on the LCD screen 4 Short the tips of the test leads together and confirm the LCD screen shows near zero 5 Connect the...

Страница 7: ...measuring range OL or OL is displayed on the LCD screen 6 Press the trigger to open the clamp jaws and place the jaws around a single conductor Easily accomplished with the use of a line splitter 7 T...

Страница 8: ...the Function Switch to OFF and remove the test leads from the instrument before trying to replace batteries 7 BATTERY REPLACEMENT WARNING Test on a known live source before use Even if the meter does...