background image

12

Measurement steps

1) Open the colorimetric tube and fill with 10ml or more of the solu

-

tion being tested, trying to avoid air bubbles. 
2) Close the colorimetric tube and wipe clean with a soft cloth.
3) Remove the shade cover and rubber seal plug from the meter 

and place the colorimetric tube in to the test cuvette, then place the 

shade cover over the colorimetric tube.

4) Power on the meter, then press “Read” to start the test.
5) Once complete, the test reading will be displayed. Press “Back” 

to return to the main menu.

Содержание 860045

Страница 1: ...1 Low Range High Precision Turbidity Meter 0 20 NTU 860045 Instruction Manual...

Страница 2: ...2 Table of Contents Overview 3 Safety Precautions 5 Features 7 Notice 9 Parts accessories 11 Measuring steps 12 Calibration steps 13 Other operation 14 Technical specifications 15 Warrenty 16...

Страница 3: ...to the ISO7027 stan dard There are two detectors 90 and 180 in the instrument The 180 detector receives the transmitted light and the 90 detector receives the scattered light The intensity of the scat...

Страница 4: ...ight interference High precision optical structure accurate and stable measure ment Built in high performance micro processing core high data pro cessing power Easy to use text on the screen indicates...

Страница 5: ...wear protective glasses if necessary If you come into contact with chem icals clean them thoroughly with water immediately and consult a doctor if necessary Please understand the measurement proce dur...

Страница 6: ...is a potentially dangerous situation If it cannot be avoided it may cause death or serious injury Warning Indicates a potentially hazardous situation that may cause minor or moderate injury Note Infor...

Страница 7: ...buttons and display screen are protected by acrylic panels 4 The test results are automatically saved and 5000 test results can be stored so that historical data can be queried 5 USB interface utiliz...

Страница 8: ...8 Features Continued 8 Support multi point calibration 9 One key Restore the settings function 10 Special carrying case dust proof and shockproof...

Страница 9: ...acing the bottle in the slot 3 Keep the white bottle intact without scratches and clean inside and outside 4 The silk screen V on the black bottle should be aligned with the on the edge of the black g...

Страница 10: ...20NTU must be calibrated 7 Please put on the shading cover before testing 8 Avoid any liquid or foreign objects entering the detection tank After use please cover the rubber sealing plug in time and p...

Страница 11: ...11 Parts accessories 1 Shading cover 2 Display Screen 3 Buttons 4 USB 5 Cuvette 6 Colorimetric Tube 7 Rubber seal plug...

Страница 12: ...ric tube and wipe clean with a soft cloth 3 Remove the shade cover and rubber seal plug from the meter and place the colorimetric tube in to the test cuvette then place the shade cover over the colori...

Страница 13: ...standard in to the col orimeter tank Place the shade cover over the colorimetric tube then press READ to start the calibration 3 Repeat step 2 for each highlighted calibration standard 4 Once all cal...

Страница 14: ...trument and dis play the drive Z10A on the computer The test records are saved in the Record folder CalibrateRecord csv is the calibration data log DetectionRecord csv is the measurement data log You...

Страница 15: ...0 01 0 1 1 0 NTU Accuracy 0 3 NTU or 8 Repeatablility 10 NTU 0 1NTU 10 NTU 1 NTU Zero Drift 0 3 FS 30 min Operating Temp 5 40 C Storage Temp 10 55 C Humidity 0 80 RH Power Lithium Battery Calibration...

Страница 16: ...battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit p...

Отзывы: