background image

- 5 - 

IV.  OPERATING INSTRUCTIONS 

 

A.

TEMPERATURE COMPENSATION 

- pH measurement. 

Note:  The meter is factory set for a solution temperature of 77°F/25°C. 

 

1.  Manual Temperature Compensation for pH Measurement 

Slide the 

LOCK SWITCH 

to the locked position and press the 

POWER 

button. The LCD flashes momentarily. 

The temperature is shown in the lower display 
(Fig.2).  Press the 

°C/°F

 button to select the 

desired unit of measure. 

Press the 

TEMP.C

 button to adjust the manual 

temperature compensation by using the 

LEFT

UP

 

and 

DOWN 

buttons. 

Press the 

TEMP.C

 button 3 times to return to the measuring mode. 

 

2.  Automatic Temperature Compensation (ATC) for pH Measurement 

Connect the pH electrode to the 

BNC SOCKET

.  Plug the ATC probe into 

the 

OPTIONAL PROBE INPUT SOCKET

.  Slide the 

LOCK SWITCH 

to the 

locked position and press the 

POWER 

button.  The LCD flashes momen-

tarily. 

Press the 

PH/MV 

button and select the pH mode.  "PH" is displayed on the 

LCD. 

Insert both probes into the solution.  The pH measurement will be automati-
cally compensated for temperature. 

 

B. PH CALIBRATION PROCEDURES  

The meter is factory calibrated to simulate the ideal pH electrode (0 mV at pH 
7.00 and 177.4 mV at PH 4, based on an ambient temperature of 77°F/25

°

C ). 

 

1.  Electronic Calibration of the Meter Only (without buffers)

 

If applicable detach the ATC probe.  Slide the 

LOCK SWITCH 

to the locked 

position and press the 

POWER 

button. The LCD flashes momentarily.  

Attach the pH probe to the 

BNC SOCKET

and press the 

PH/MV 

button to 

select the pH mode ("PH" is displayed on the LCD). 

To adjust the pH 4 default value, press the 

TEMP.C 

button twice.  The upper display will show "4.00"

 

and the lower display will show the pH 4 default 
calibration value (Fig.3).  Use the 

UP 

and 

DOWN 

buttons to adjust the value. 

 

Fig. 3

 

         PH 
         4.00 
             04.03

 

Fig. 2 

        PH 
 
              026.1 °C

 

See Fig.1, Pg.4 for inputs & button locations

 

Содержание 850081

Страница 1: ...Water Quality Meter 850081 Instruction Manual SPER SCIENTIFIC LTD ...

Страница 2: ...IBRATION PROCEDURE 5 1 Electronic Calibration without buffers 5 6 2 Chemical Calibration with buffers 6 7 3 Meters that are Out of Calibration by more than 1 pH of the Calibration Solution 7 C MEASUREMENT PROCEDURES 7 1 Temperature 7 2 PH 8 3 mV 8 4 Hold 8 5 Maximum Minimum 8 9 D AUTOMATIC SHUT OFF 9 E OVERLOAD INDICATOR 9 F BATTERY REPLACEMENT 9 G RS 232 PC SERIAL INTERFACE 9 10 V SPECIFICATIONS ...

Страница 3: ...temperature in user selectable ºC ºF and all measurements are automatically temperature compensated The RS 232 port enables communication with a computer Also features automatic shut off hold maximum minimum fold out easel back and a tripod screw Comes with 9V battery and a hard shell foam lined carrying case with room for one or two probes 1 Display 7 C Down CAL Button 12 Opt Probe Input Socket 2...

Страница 4: ...es on the front panel Automatic temperature compensation by optional ACT probe Measurement Power Management OR Auto Power Off not activated during Memory Record Selection Manual Power Off Power On Single point calibration pH7 CAL Set manual temp values or attach ATC probe Two points calibration pH7 CAL PH4 pH10 SLOPE Calibration III Procedures Overview ...

Страница 5: ...ER button The LCD flashes momen tarily Press the PH MV button and select the pH mode PH is displayed on the LCD Insert both probes into the solution The pH measurement will be automati cally compensated for temperature B PH CALIBRATION PROCEDURES The meter is factory calibrated to simulate the ideal pH electrode 0 mV at pH 7 00 and 177 4 mV at PH 4 based on an ambient temperature of 77 F 25 C 1 El...

Страница 6: ...e are needed to perform the following procedures Two Point Calibration Connect the PH electrode to the BNC SOCKET Ensure that the temperature setting is adjusted to the temperature of the pH buffer solution See TEMPERATURE COMPENSATION on page 5 Press the PH MV button and select the pH mode PH is displayed on the LCD PH 7 Calibration Rinse the electrode with distilled water and immerse the electro...

Страница 7: ...H 10 when the measurement is pH 11 or pH 9 Connect the pH electrode to the PH BNC INPUT SOCKET Slide the LOCK SWITCH to the locked position and press the POWER button The LCD flashes momentarily Press the PH MV button and select the pH mode PH is displayed on the LCD Set the Manual Temperature Compensation value to 77 F 25 page 5 Rinse the electrode with distilled water and place it into a standar...

Страница 8: ...e the electrode into the solution The mV measurement is displayed Rinse the electrode with distilled water after each use 4 Hold During the measuring procedure press the HOLD button to freeze the display The LCD will display the word HOLD Press the HOLD button again to exit 5 Maximum Minimum To record the maximum and minimum readings Press the REC button once REC appears on the LCD Press the REC b...

Страница 9: ...low battery icon is displayed in the left corner of LCD In spec measurements may be made for several hours after the low battery indicator appears Slide the battery cover away from the instrument remove the battery and replace with a 9V battery alkaline or heavy duty type Close the battery cover G RS232 PC SERIAL INTERFACE The instrument features RS232 output via a 3 5 mm terminal The signal outpu...

Страница 10: ...V 18 mS 14 0 6F1 06 mg L 07 D13 The upper display data 1 The lower display data 2 D14 4 D15 Start Word RS232 FORMAT 9600 N 8 1 V SPECIFICATIONS Circuit Custom one chip microprocessor LSI circuit Display 51 mm x 32 mm dual function LCD display 15mm 0 6 digit size Input Impedance 1012 ohm Temperature Compensation for pH measurement Manual 0 100 C adjusted by the C F button Automatic ATC With optiona...

Страница 11: ...ftware 840057 RS232 Cable PROBES Unit of Msrmt Range Resolution Accuracy pH Standard BNC connector 840016 pH 0 14 0 01 0 02PH 2d pH Spear Tip Standard BNC connector 840049 pH 0 14 0 01 0 02PH 2d ATC Temp Probe 850082 C 0 65 0 1 0 8 F 32 149 1 5 Conductivity Probe 850083 mS 0 2 1 999 0 001 3 fs 1d mS 2 19 99 0 01 ºC 0 60 0 1 0 8 ºF 32 140 1 5 Conductivity TDS Probe 850084 mS 0 2 2 000 0 001 3 fs 1d...

Страница 12: ...ivalent Sper Scientific product will be substituted if available This warranty does not cover probes batteries or damage resulting from accident misuse or abuse of the product In order to obtain warranty service simply ship the unit postage prepaid to SPER SCIENTIFIC LTD 7720 East Redfield Suite 7 Scottsdale Arizona 85260 480 948 4448 www sperscientific com Please Note The defective unit must be a...

Отзывы: