Spencer 170 NXT Скачать руководство пользователя страница 19

19

 

12.

 

CLEANING AND MAINTENANCE 

12.1

 

Cleaning 

Failure to carry out the correct cleaning routine could increase the risk of cross infection, due to presence of body fluids and/or residuals.  
The operator must always wear adequate personal protection such as gloves and mask, etc. during all checking and cleaning procedures. 
 

 

The described operations must be performed after each use of Spencer 170. 

 

Turn off the device 

 

Isolate it from the power supply (if connected) 

 

The cleaning of the external parts of the device can be carried out using suitable disinfectants for surfaces as described in this table: 
 

USABLE PRODUCTS 

NOT USABLE PRODUCTS 

Disinfectants with aldehydes 

Compounds that release halogens 

Disinfectants with alcohol 

Strong organic acids 

Quaternary ammonic compounds 

Compounds that release oxygen 

 

Trichloroethylene 

 

 

Be careful to ensure that no kind of solution enters inside the oxygen socket.. 

 

If is used a disposable patient circuit, replace it.  

 

If is used a reusable patient circuit, disassemble every part and sterilize according to a validated procedure. 

 

Similarly, replace or sterilize the mask.  

 

After confirming the ventilator and all accessories are completely dry, is possible to reconnect the device to the power supply.  

The use of high pressure water is prohibited

,

 because it should penetrate into the device causing risks of corrosion of components 

and risks of electric leakage and short circuits. During every cleaning procedure, check that were not used any kind of lubricants.  

 

12.2

 

ORDINARY maintenance 

Establish  a  maintenance  program  and  periodic  testing  routine,  identifying  an  employee  responsible  for  this.  The  person  to  whom  the 
ordinary maintenance of the device is entrusted must ensure the basic requirements foreseen by the manufacturer in following paragraphs 
are inspected. 
All maintenance and periodic servicing activities must be registered and kept together with the servicing reports. These documents have to 
be kept for a period of 10 years after the  disposal of the device itself. This  register will be made available to the competent authorities 
and/or manufacturer if requested. 

 

To guarantee the traceability of the products and to protect the procedures of maintenance and assistence of Yours devices, Spencer has 
made available the ASSTEC portal 

http://service.spencer.it/asstec/login.aspx

which will allow you to view the data of the products owned 

or on the market, to monitor and update the plans of periodic reviews, to view and manage extraordinary maintenances. 

 

Routine maintenance of the device must be carried out by operators in possession of specific qualifications, trained and experienced in the 
use and maintenance of the device. 

The operator must always wear adequate personal protection such as gloves and mask, etc. during all checking and cleaning procedures. 

 

Checks to be carried out before and after each use and at deadline indicated above, are as follows: 

 

General functionality of the device 

 

Cleanliness of the device (remember that the failure of cleaning may cause the risk of cross infections) 

 

Correct fixation of all nuts, bolts and screws 

 

No structural part is deformed or compromised 

 

Proper operation of the pressure limiting adjustment controls. With the device turned on, occlude the patient output and verify that 
the manovacuometer shows the pressure value set within the tolerances described in the technical data. Verify proper functionality of 
the frequency knob setting a value and measuring the time between two mandatory ventilations. The time in seconds between two 
ventilations is equal to 60/Frequency. 

 

12.2.1

 

9V alarm battery check

 

The 9 V backup battery must be regularly checked or replaced depending on its use. Using a tester ensure that the voltage on the battery is 
not lower than 40%. The correct polarity when connecting the 9 V battery is guaranteed by one-directional connector.  

The alarm battery must be replaced once a year even if in good conditions. 

 

12.2.2

 

Internal battery status check 

The device is powered by a rechargeable battery. A new battery is never fully charged. It will be fully charged only after being charged for 7 
consecutive  hours  before  the  first  use  of  the  device.  The  battery  recharges  automatically  when  the  device  is  connected  to  the  external 
power source. 

If the device is not used for a long time, the battery must be recharged. 
Do not keep the  unit in charge  beyond the time specified. Fully discharge the battery and / or maintain the equipment  continuously 
powered from external source, results in a significant reduction of the operating life of the battery. 
 
The internal battery must be replaced once a year even if in good conditions. 
All batteries should be removed if the device needs storage. The absence of the internal battery should be indicated on te device (for 
example with a label) so that the device will not be accidentally used without the internal power supply. To limit the battery aging, it is 

Содержание 170 NXT

Страница 1: ...d may vary slightly from the actual device Spencer Italia S r l assumes no responsibility for any errors contained herein or for damage accidents or consequences connected with the supply performance or use of this manual Prima emissione 1998 Rev 3 21 05 2019 SPENCER ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy www spencer it support spencer it e mail info spencer it service service...

Страница 2: ...ESHOOTING 23 13 1 Modalità di rientro per riparazione 24 13 1 How to return for servicing 24 14 ACCESSORIES 24 15 SPARE PARTS 24 16 DEMOLITION 25 Avvertenza Le informazioni contenute in questo documento sono soggette a modifica senza preavviso e sono da intendersi come impegno da parte della Spencer Italia S r l con riserva di modifiche I prodotti Spencer vengono esportati in molti paesi nei quali...

Страница 3: ...sts Regulation 10 Uniform provisions concerning the approval of vehicles with regard to electromagnetic compatibility 4 INTRODUCTION 4 1 Use of the Manual This manual is intended to provide to the health care operator all the necessary information for its safe and appropriate use as well as adequate maintenance of the device Note this Manual is an integral part of the device therefore it must be k...

Страница 4: ...ep in a cool and dry place IP34 Protection of enclosures for electrical devices First digit protection against ingress of particulate greater than 2 5 mm diameter Second digit Protected against splashing water from all directions Warning for the correct disposal of the product according to the European Directive 2012 19 UE External power supply Internal power supply I ON O OFF AUDIO PAUSED Non ion...

Страница 5: ... will be made available to the competent Authorities and or manufacturer if requested In the absence of such documentation sanctions will be applied Do not allow any untrained person to help during the use of the product because they could cause damage to themselves or to others Note Spencer Italia S r l is always at your disposal to organise product training Installers training The installer of t...

Страница 6: ...end of life of the product and must be made available to the competent authorities and or the Manufacturer if requested The cleaning scheduled for reusable products must be performed in accordance to the directions provided by the Manufacturer in the User Manual in order to avoid the risk of cross infections due to the presence of secretions and or residuals The device and all of its components if...

Страница 7: ...s or wrong measurement of ventilation parameters Portable RF communications equipments should be used no closer than 30 cm from any part of the device including cables Otherwise the performance of the device could be compromised In order to maintain the essential performance and basic safety of the device is essential to check before and after each use the integrity of the electrical connections a...

Страница 8: ...erapy For the use of the ventilator a specialized doctor must be present The doctor will be able to determine if the technical specifications of the device makes it suitable to be used on a specific patient and will be the sole responsible for the definition and setting of the ventilation parameters Do not use the device in presence of inflammable substances and anesthetics The device is not inten...

Страница 9: ...leakage decreasing the effectiveness of the ventilation therapy or leading to improper operation of alarms The artificial ventilation can have side effects In order to identify the hazards associated with the use of the device related to the clinical conditions of the patient its essential the presence of an expert doctor who can evaluate the actual benefits provided by the artificial ventilation ...

Страница 10: ...tion according to IEC 60601 1 Classe II if externally powered Internally powered if operating with internal battery Classification applied part Breathing tube filter connector valve mask BF Enclosure protection according to IEC 60529 IP44 Protected against ingress of particles with diameter greater than 12mm Protected against splashing water from all directions 8 2 Components N Description Materia...

Страница 11: ...y would be voided and Spencer Italia will not have any liability related to the functionality and use of the product The pneumatic circuit can be modified or implemented without prior notice 9 INSTALLATION AND START UP 9 1 Installation Warning The device is shipped with unplugged battery Connect the buttery before the start up The installation of the device is a critical step to ensure a proper op...

Страница 12: ... near any part of the appliance including cables etc and should be kept at a distance never less than the recommended and calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 0 583x P d 1 2 x P da 80Mhz a 800MHz d 2 3 x P da 800Mhz a 2 7 GHz where P is the maximum rated power output of the transmitter in watts W according to the transmitter ...

Страница 13: ...and reflection from structures objects and people Guide and Manufacturer s Declaration Immunity to proximity fields from RF wireless communications equipment Test frequency MHz Modulation Immunity level V m 385 Pulse modulation 1 at 18Hz 27 450 FM 2 5Hz deviation 1kHz sine 28 710 Pulse modulation 1 at 217Hz 9 745 Pulse modulation 1 at 217Hz 9 780 Pulse modulation 1 at 217Hz 9 810 Pulse modulation ...

Страница 14: ...ion on a flat floor Drill two holes on the surface for the screws Choose screws of appropriate length according to the installation surface Tighten the screws until a safe fix is reached Make sure the screws are fully seated If they are not perfectly coupled with the surfaces is necessary to choose shorter ones or if the distance between the screw and the surface is reduced apply washers 9 2 Start...

Страница 15: ... AIR MIX selector Allows to choose whether to use the medical gas delivered by the medical gas supply system at 100 NO AIR MIX or at 60 AIR MIX mixing it with air taken from the environment in which is placed the device F ON OFF switch Turn on the device by pressing the I button turn it off by pressing the O button G Alarm LED section Provide visual feedback regarding the activated alarms or their...

Страница 16: ...his LED To disconnect the device from the pneumatic supply push the outer plastic part of the socket in axial direction while pulling out the probe 11 2 Turn on the device Turn on the device by pressing on the button identified with the symbol I The device will perform an autodiagnostic test by turning on all LEDs for about 1 second and emitting 3 short sounds When this test is concluded the devic...

Страница 17: ...ously If connected to the external power supply and the battery is not fully charged the charging will automatically start and the white led next to the symbol will light Flashing of the white led next to the symbol identifies a particular condition that do not allow to charge the battery See par 11 6 The real pressure in the patient circuit is always shown on the manovacuometer placed on the fron...

Страница 18: ...s of activation 1 The battery has a remaining autonomy of about 10 minutes Action to take 1 Connect the device to the external power source as soon as possible Technical medium priority alarms LOW BATT Alarm specification 3 sequences of sound pulses spaced 2 5 seconds and dedicated yellow led flashing one time every two seconds Reason of activation The primary battery has reached a charge level th...

Страница 19: ...the data of the products owned or on the market to monitor and update the plans of periodic reviews to view and manage extraordinary maintenances Routine maintenance of the device must be carried out by operators in possession of specific qualifications trained and experienced in the use and maintenance of the device The operator must always wear adequate personal protection such as gloves and mas...

Страница 20: ...must be replaced every year 12 2 4 Battery replacement The annual maintenance of the device requires among other activities provided by the manufacturer the replacement of all batteries If is necessary to replace the batteries before the annual revision follow these instructions Verify that the device is turned off and disconnected from the power supply Using a coin or a screwdriver open the batte...

Страница 21: ...uito descritte devono essere effettuate prima di ogni utilizzo dell apparecchio To check Required result RESPIRATORY SYSTEM Corrugated Tube Non rebreathing valve PEEP valve if present Face mask Ventilation test Disposable filter Connection All components have to be in good conditions and correctly connected The device or its components must be correctly cleaned or replaced ELECTRIC POWER SUPPLY Tu...

Страница 22: ... Spencer Italia S r l and the end user to keep a log book regarding the operations carried out on the device 12 4 EXTRAORDINARY maintenance Only the manufacturer or centres with written authorisation are authorised to complete any special servicing operations For any operations that are not carried out directly by the manufacturer but by an authorised centre we have to underline that a report rega...

Страница 23: ...e battery The fuse is blown Replace the fuse as described in the manual The device is not working Low voltage has caused the block of the microprocessor Verify the voltage of the power supply Turn off the device and turn on again If the problem persists contact a Service center Turning on the device it does not start the controlled ventilation Failure to electrical or pneumatic supplies or a failu...

Страница 24: ...d Request to specify the defect in order to carry out the repair in the shortest possible time It therefore requires to carefully read the instructions to avoid compromising the device with inappropriate use It requires you to specify the kind of fault to give way to the technical Spencer Italy S r l to judge whether the fault falls into the category of warranty 14 ACCESSORIES Standard equipment E...

Страница 25: ...waste Can be taken to special recycling centers provided by local government or return it to the dealer on purchase of a new device of the same type and used for the same functions Dispose of the product separately avoids possible negative consequences for the environment and human health resulting from inappropriate disposal and allows to recover the materials in order to obtain significant savin...