background image

EPM/SMS MANUAL 

16 

118026-001  REV K 

Chapter 5 

O

PTIONS

 

He  options  available  for  this  power  supply  are 
described  in  this  section.    Interface  diagrams  are 
shown  where  required.    Options  are  specified  by 

including  the  option  code  in  the  model  number  as 
described in Section 1.4. 

5.1  Floating Ground   

        FG 

The  floating  ground  option  allows  isolation  of  the  power 
supply  common  from  chassis  ground.    A  typical 
application  for  FG  is  for  connection  of  an external meter 
or  circuit  into  the  FG  signal  point.    This  circuit  can 
monitor  actual  current  flow  at  a  low  voltage  level.    See 
Figure  5.1  for  typical  connection  to  the  floating  ground 
terminal. 

IMPORTANT 

The power supply common must be 

connected to chassis ground through a 

low impedance circuit.  The power 

supply common is clamped to chassis 

ground using a power zener diode.  This 

diode is not intended to clamp under 

continuous operation.  On higher 

current units the power dissipation 

within the zener diode can be excessive 

if allowed to clamp. 

5.2  24Vdc  

 

 

 

24 

EPM/SMS  series  power  supplies  with  the  24Vdc  input 
option will operate from an input voltage of 24Vdc±10% 

5.3  No Slow Start 

 

      NSS 

The  no  slow  start option causes the output voltage of the 
power  supply  to  rise  (within  500  msec)  to  the  rated 
voltage upon Power Up. 

5.4  Non-Standard Slow Start 
 

 

 

 

 

    SS(x) 

The non-standard slow start option allows the gradual rise 
time  of  the  output  voltage  to  be  different  from  the 
standard  of  six  seconds.    To  order  the  option,  place  the 
time  desired  in  seconds  after  the  suffix  letter;  i.e.  SS(10) 
denotes a 10 second rise time. 

5.5  Extra Length Output Cable 
 

 

 

 

 

    LL(ft) 

Standard output cable is 18” long.  Other lengths may be 
specified. 

5.6  Custom Designed Models  X(#) 

Units  built  to  customer  specifications  are  assigned  an  X 
number  be  the  factory.    If  this  unit  is  an  X  model, 
specification  control  sheet  is  added  at  the  end  of  this 
instruction manual. 

Spellman welcomes the opportunity to tailor its units to fit 
your  requirements  or  to  develop  new  products  for  your 
applications.    Contact  the  Spellman  Sales  Department 
with your needs. 

 

Содержание EPM Series

Страница 1: ...SERIAL DATE SPELLMAN HIGH VOLTAGE ELECTRONICS CORPORATION 475 Wireless Blvd Hauppauge New York 11788 1 631 630 3000 FAX 1 631 435 1620 E mail sales spellmanhv com Website www spellmanhv com EPM SMS MANUAL 118027 001 Rev G EPM SMS SERIES ...

Страница 2: ...spannungsnetzteiles benötigen ca 5 Minuten um sich zu entladen Erden Sie sich nicht und arbeiten Sie nicht in feuchter oder nasser Umgebung Notwendige Reparaturen können es erforderlich machen den Gehäusedeckel während des Betriebes zu entfernen Reparaturen dürfen nur von qualifiziertem eingewiesenem Personal ausgeführt werden WARNING im folgenden Text weist auf gefährliche Operationen hin die zu ...

Страница 3: ...de l alimentation Ne vous mettez pas à la masse ou ne travaillez pas sous conditions mouillées ou humides La maintenance peut nécessiter l enlèvement du couvercle lorsque l alimentation est encore allumée Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques Dans le manuel les notes marquées WARNING attire l attention sur les risques lors de la m...

Страница 4: ...carica della capacità interna dell alimentatore ad alta tensione Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d umidità SICUREZZA NELLA MANUTENZIONE Manutenzione potrebbe essere richiesta rimuovendo la copertura con apparecchio acceso La manutenzione deve essere svolta da personale qualificato coscio dei rischi elettrici Attenzione alle AVVERTENZE contenute nel ...

Страница 5: ...allation 3 2 3 Installation Instructions 6 3 OPERATING INSTRUCTIONS 3 1 Operation 7 3 2 Standard Features 8 4 PRINCIPLES OF OPERATION 4 1 High Frequency Inverter 15 4 2 High Voltage Circuits 15 4 3 Control Circuits 15 4 4 Options 15 5 OPTIONS 5 1 Floating Ground 16 5 2 24Vdc 16 5 3 No Slow Start 16 5 4 Non Standard Slow Start 16 5 5 Extra Length Output Cable 16 5 6 Custom Designed Models 16 6 MAIN...

Страница 6: ... 1 SMS Typical Operating Set Up 9 3 2 EPM Typical Operating Set Up 9 3 2A SMS Remote Programming via Voltage Source 10 3 2B SMS Remote Programming via External Resistance 10 3 3 SMS Remote Monitoring 11 3 4A SMS HV ON OFF Enable Low Control 11 3 4B SMS HV ON OFF Enable Hi Control 12 3 5A EPM Remote Programming via Voltage Source 12 3 5B EPM Remote Programming via External Resistance 13 3 6 EPM Rem...

Страница 7: ...M SMS series modules Consult factory about the warranty for custom modules 1 2 EPM SMS Specifications Output Control Voltage and current are externally programmable over the entire range from zero to maximum rating via 0 10VDC reference or potentiometer Other scale factors for control and monitoring can be provided i e 1 volt per 10KV 1 volt per 10mA etc Input Voltage 28Vdc 10 28Vdc 10 optional Co...

Страница 8: ...age and current via an external voltage source Remote Monitor Allows remote monitoring of the output voltage and current High Voltage Enable Disable Allows remote ON OFF control of the high voltage 10VDC Reference A 10VDC reference is provided for remote programming via a potentiometer or resistive divider 1 4 Options The options available are listed in Table 1 1 See section 5 for more information...

Страница 9: ...s and components are covered by warranty Custom and special order models with an X suffix in the model number are also covered by warranty 2 2 Mechanical Installation The EPM SMS series module power supplies are designed for installation into existing or newly developed OEM equipment The power supply can also easily fit into bench top applications or test set requirements Standard unit dimensions ...

Страница 10: ...mote adjustment of the output voltage and current via an external voltage source By adjusting the voltage source from 0V to 10V the desired output may be selected Remote Monitor Test points pins 4 and 5 of the input connector monitor voltage and current outputs The output polarity is positive from 0V to 10V equal to 0 to 100 of the output 0 5 Output impedance is 1k Ohms High Voltage Enable Inhibit...

Страница 11: ...ment of the output voltage and current via an external voltage source By adjusting the voltage source from 0V to 10V the desired output may be selected Remote Monitor Test points pins 4 and 5 of the input connector monitor voltage and current outputs The output polarity is positive from 0V to 10V equal to 0 to 100 of the output 0 5 Output impedance is 1k Ohms High Voltage Enable Inhibit Control si...

Страница 12: ...60 30kV 2mA 60W SMS40X60 40kV 1 5mA 60W SMS50X60 50kV 1 2mA 60W SMS60X60 60kV 1mA 60W Table 2 1 Output Voltage and Current Ratings for 24Vdc or 28Vdc 5A Models Dieses Geraet ist zugelassen gemaess EN60950 1993 AM1 1993 AM2 1993 and EN61010 1 1993 Im falle das diesses Geraet in Informationstechnik einrichtung eingegaut werden soll die installation muss gemaess den obengenannten Norn erfolgen Das Ga...

Страница 13: ...ng a separate external ground at the load is not recommended D Options Note See section 5 for hook up and operating instructions for the options on your unit Custom models may also require set up changes E Hook up Connect control and monitoring connections as described in this manual F For initial turn on program the voltage and current for zero output Connect the enable disable signal to disable ...

Страница 14: ... input connector for current programming Programming signals should be referenced to Pin 9 of the control input connector signal ground By adjusting the voltage source from 0 volts zero output to 10 volts full rated output the desired output can be selected See Figure 3 2A and 3 5B for wiring diagram and specifications An alternate method of controlling the output remotely is by using external res...

Страница 15: ...EPM SMS MANUAL 9 118026 001 REV K Figure 3 1 SMS Typical Operating Set Up Figure 3 2 EPM Typical Operating Set up ...

Страница 16: ...10 118026 001 REV K Figure 3 2A SMS Remote Programming Via Voltage Source ____________________________________________________________________________________ Figure 3 2B SMS Remote Programming Via External Resistance ...

Страница 17: ...EPM SMS MANUAL 11 118026 001 REV K Figure 3 3 SMS Remote Monitoring ____________________________________________________________________________________ Figure 3 4A SMS HV ON OFF Enable Low Control ...

Страница 18: ...MANUAL 12 118026 001 REV K Figure 3 4B SMS HV ON OFF Enable Hi Control ____________________________________________________________________________________ Figure 3 5A EPM Remote Programming Via Voltage Source ...

Страница 19: ...MS MANUAL 13 118026 001 REV K Figure 3 5B EPM Remote Programming Via External Resistance ____________________________________________________________________________________ Figure 3 6 EPM Remote Monitoring ...

Страница 20: ... SMS MANUAL 14 118026 001 REV K Figure 3 7A EPM HV ON OFF Enable Low Control ____________________________________________________________________________________ Figure 3 7B EPM HV ON OFF Enable Hi Control ...

Страница 21: ...done by the low voltage control circuits The resonant circuitry consists of a resonant inductor and step up transformer 4 2 High Voltage Circuits The high voltage transformer is a step up type The secondary of the high voltage transformer is connected to the high voltage rectifier circuit The rectifier circuit will vary depending upon the rated output voltage For lower output voltage a full bridge...

Страница 22: ...ion within the zener diode can be excessive if allowed to clamp 5 2 24Vdc 24 EPM SMS series power supplies with the 24Vdc input option will operate from an input voltage of 24Vdc 10 5 3 No Slow Start NSS The no slow start option causes the output voltage of the power supply to rise within 500 msec to the rated voltage upon Power Up 5 4 Non Standard Slow Start SS x The non standard slow start optio...

Страница 23: ...EPM SMS MANUAL 17 118026 001 REV K Figure 5 1 Floating Ground Interface Circuit ...

Страница 24: ... can be obtained from the Spellman Customer Service Department Test equipment including an oscilloscope a high impedance voltmeter and a high voltage divider such as the Spellman HVD 100 or HVD 200 is needed for performance tests All test components must be rated for operating voltage 6 3 High Voltage Dividers High voltage dividers for precise measurements of output voltage with an accuracy up to ...

Страница 25: ... 7 2 Correspondence and Ordering Spare Parts Each power supply has an identification label on the rear of the chassis that bears its model and serial number When requesting engineering or applications information please state the model and serial number of the power supply If specific components or circuit sections are involved in the inquiry it is helpful to indicate the component symbol number s...

Страница 26: ...o the customer The Customer shall pay for shipping to and from Spellman A preliminary estimate for repairs will be given by phone by Customer Service A purchase order for this amount is requested upon issuance of the RMA Number A more detailed estimate will be made when the power supply is received at the Spellman Repair Center In the event that repair work is extensive Spellman will call to seek ...

Страница 27: ...To obtain information on Spellman s product warranty please visit our website at http www spellmanhv com en About Warranty aspx ...

Отзывы: