The  data  is  stored  in  comma-delimited  text  files.    These 
files can be opened with text-editing software (e.g. Micro-
soft Word) or spreadsheet software (e.g. Excel). 

 

The first two lines of the data file give the logger’s name 
and  serial  number.    The  third  line  indicates  that  latitude 
and longitude are referenced to the 1984 World Geodetic 
Survey datum.  The fourth line shows the cone index unit 
system.  The fifth line shows the column headings for the 
rest of the data file.  Each row of data represents a meas-
urement profile. 

 

Logging sessions are started and completed by turning the 
meter  on  and  off.    The  start  of  a  logging  session  is  indi-
cated  by  the  data  line  “Logger  Started”.    If  a  GPS  signal 
was found at the start of a logger session a time stamp is 
included on the “Logger Started” line.   

 

 

Sample  data  showing  results  of  data  collected  with  and 
without GPS activated.  Note: GPS signal not found when 
recording first four sample numbers

Data Management 

Data Management 

Data Management 

Data Management     

  9 

T h e  

c o m p u t e r 

Communications Port to 
which  the  PC-3.5  serial 
cable is connected can be 
identified  by  using  a 
paper clip. 

1.  Disconnect  the  serial 
cable  from the meter. 

2.  To  bring  up  the  Port  Selection  screen,  click  on  the 
Com  Port  Button,  select  the  com  port  to  be  tested  and 
click  the  Port  Test  button.    Click  the  Test  Port  Now 
button.  If  the  message  “Connection  OK”  is  displayed, 
another  device  (such  as  a  modem)  is  probably  connected 
to that port. If the message “No Connection” is displayed, 
this port may be the one connected to your serial cable and 
you can proceed to the next step. 

3. Place a paperclip on the end of the serial pin so that it 
touches both the tip of the pin and the metal area between 
the two black rings.   Again click on the Test  Port  Now  
button.    If  the  message  “Connection  OK”  now  appears, 
this is the com port connected to your serial cable.   

paper clip 
or wire 

Identifying the 

Identifying the 

Identifying the 

Identifying the     

Correct Com Port

Correct Com Port

Correct Com Port

Correct Com Port    

 

NOTE:  The  meter  does  not  short-circuit  the  serial  pin.  
Therefore,  when  the  Test  Port  Now  button  is  clicked 
while  the  meter  is  connected,  the  “No  Connection” 
message will be displayed. 

Содержание 6110FS

Страница 1: ...East Plainfield IL 60585 800 248 8873 or 815 436 4440 FAX 815 436 4460 E Mail info specmeters com www specmeters com Technologies Inc Spectrum Catalog 6110FS User s manual SC 900 Soil Compaction Meter Technologies Inc Spectrum ...

Страница 2: ...anagement 20 Meter Dimensions 21 Specifications 22 Appendix 1 Time Zone Corrections 23 Appendix 2 When to Replace the Cone 24 Warranty 26 This manual will familiarize you with the features and operation of your new Field ScoutTM SC 900 Soil Compaction Meter Please read this manual thoroughly before using your instrument For customer support or to place an order call Spectrum Technologies Inc 800 2...

Страница 3: ...loss of the package by any shipping company 3 Thank you for purchasing the Field ScoutTM SC 900 Soil Compaction Meter This manual describes the features and operation of the meter Soil compaction prevents moisture penetration reduces fertilizer and chemical uptake and hinders plant root growth In some cases yield losses can run as high as 30 due to compaction This new electronic soil compac tion m...

Страница 4: ...ion depths have been determined a much more effective tillage system can be employed The unit works as an efficient diagnostic tool as well as a management unit after the tillage operation has been performed One can determine how deep the soil has actually been worked and if the tillage operation has solved the compaction problems 25 When you are doing comparative analysis of the readings in a fie...

Страница 5: ...ld of view ex tends about 30o from the shaft The sensor meas ures distance by emit ting a sound wave and measuring the travel time of the first reflected wave Usually this first reflection will come from the ground just ad jacent to the shaft However if an object with a surface oriented toward the sensor is within the field of view an erroneous depth measurement may be made This can lead to depth ...

Страница 6: ...Logger 75 Full No GPS Found Sample meter power up screens with datalogger enabled left screen indicates GPS signal was found 0010 PSI 00 IN Logger OFF If when powered up the screen indicates the data logger is not activated see above screen no data will be stored by the meter The data logger can be activated through the SC 900 software see Meter Settings pg 14 23 Appendix 1 Time zone corrections T...

Страница 7: ...ration cont Meter operation cont The START button readies the meter and datalogger to take a profile meas urement It must be pressed before every attempted profile measurement START After measuring a profile pressing the REVIEW button allows the user to see the cone index value at each depth As the button is pressed the display screen will scroll sequen tially through the entire profile REVIEW Imm...

Страница 8: ...ce ca ble This cable connects to the 9 pin serial port of your computer and to the meter s computer port The meter s configuration can be modified by clicking on the Meter Settings button see Meter Settings p 14 The Com Port Meter Type Download Clear Memory and Me ter Settings buttons are explained in the Main Toolbar section p 12 SC 900 data port 21 Meter dimensions The following are the dimensio...

Страница 9: ...ent 9 T h e c o m p u t e r Communications Port to which the PC 3 5 serial cable is connected can be identified by using a paper clip 1 Disconnect the serial cable from the meter 2 To bring up the Port Selection screen click on the Com Port Button select the com port to be tested and click the Port Test button Click the Test Port Now button If the message Connection OK is displayed another device ...

Страница 10: ...al check that it has the following settings Data bits 8 Stop bits 1 Baud rate 4800 bps Parity None Timing 1 second GGA data string Spectrum GPS DGPS Cable SC900 Meter GPS Unit GPS computer interface cable Connecting to a GPS Unit 19 Warning Messages Cont This error message indi cates that the meter missed a depth reading This can be caused by a momentary but rapid insertion of the probe into the s...

Страница 11: ... cannot compute the cone index You may see this message more often when taking readings in areas with non compacted zones beneath strongly compacted zones Start Over Insertion too fast 11 Calibration Calibration Calibration Calibration Battery Change Battery Change Battery Change Battery Change Calibration The reading of the meter s load cell can be recalibrated with the following procedure Note D...

Страница 12: ...nd evenly so that there is not any side stress on the shaft If an error message is generated see Warning Messages pg 18 start over by removing probe and pushing the Start button 5 Remove probe gently If the profile was success fully measured the following message will be dis played 6 You can review compaction measurements increments by pushing the Review button You can also delete the profile by p...

Страница 13: ...en soil surface you may need to use a target to ensure reliable performance see Depth Sensor p 5 It is advisable to smooth the soil surface as much as possible before inserting the probe As the probe is inserted into the soil the 0010 PSI 00 IN N 1 Profile Information Screen 13 Download To download data from the internal data logger turn the meter off and connect the gray serial cable to the RS 23...

Страница 14: ...on If a signal is found position data will be stored along with the soil compaction data If no GPS signal is available when the logger is turned on the logger will no longer look for one when measuring and recording soil moisture data If the second box is 15 checked the logger will store the GPS value only if it has been differentially corrected If the differential correction is not found only the...

Отзывы: