Spectrum 6110FS Скачать руководство пользователя страница 5

24 

Appendix 1 

Determining When to 

Replace Cone 

Repeated soil sampling will wear down the metal cone.  
Periodically  measure  the  diameter  of  the  cone  using  a 
dial  caliper  or  micrometer  that  allows  you  0.001-inch 
measurements.   
 
Place  the  caliper/micrometer  on  the  shoulder  of  the 
cone.  Figure 1 shows the proper placement and Figure 
2 the layout of the shoulder. 

Figure 1

Shoulder

Figure 2

The  determination  as  to when  to  replace  the  cone  will 
depend on how the user is applying the meter.   
 
If  you are using the meter to get absolute compaction 
readings, replace the cone when there is a “wear factor” 
of  3%.    Three  percent  wear  affects  the  cone  index 
measurements  by  5%.    A  new  cone  has  a  diameter  of 
0.505“, see Figure 3.  When the cone diameter reaches 
0.490“ (3% wear), replace the cone. 

  5 

An  ultrasonic  sensor  lo-
cated  at  the  base  of  the 
meter is used to measure 
the depth of penetration.  
The  field  of  view  ex-
tends about 30

o

 from the 

shaft.  The sensor meas-
ures  distance  by  emit-
ting  a  sound  wave  and 
measuring 

the 

travel 

time of the first reflected 
wave.  Usually, this first 
reflection 

will 

come 

from the ground just ad-
jacent  to  the  shaft.  
However,  if  an  object 
with  a  surface  oriented 
toward  the  sensor  is 
within  the  field  of  view, 
an 

erroneous 

depth 

measurement  may  be 
made.    This  can  lead  to 
depth error messages (see Warning  Messages  p. 18).  If 
possible,  try  to  remove  the  object  and/or  smooth  the  sur-
face.    Repositioning  your  feet  and  knees  may  also  elimi-
nate these errors.  If the problem persists, use a flat target 
surface with a hole slightly large than the diameter of the 
cone  to  test  the  depth  sensor  and,  if  necessary,  to  take 
measurements.    This  target  should  be  about  1  foot  wide 
and can be made out of cardboard or something sturdier if 
more durability is required.  To test the sensor, lay the tar-
get  on  the  ground  and  place  the  shaft  in  the  hole.    Have 
another person slowly raise the target while you verify the 
meter is incrementing the depth. 

 

Depth 
Sensor 

Depth Sensor  

Содержание 6110FS

Страница 1: ...East Plainfield IL 60585 800 248 8873 or 815 436 4440 FAX 815 436 4460 E Mail info specmeters com www specmeters com Technologies Inc Spectrum Catalog 6110FS User s manual SC 900 Soil Compaction Meter Technologies Inc Spectrum ...

Страница 2: ...anagement 20 Meter Dimensions 21 Specifications 22 Appendix 1 Time Zone Corrections 23 Appendix 2 When to Replace the Cone 24 Warranty 26 This manual will familiarize you with the features and operation of your new Field ScoutTM SC 900 Soil Compaction Meter Please read this manual thoroughly before using your instrument For customer support or to place an order call Spectrum Technologies Inc 800 2...

Страница 3: ...loss of the package by any shipping company 3 Thank you for purchasing the Field ScoutTM SC 900 Soil Compaction Meter This manual describes the features and operation of the meter Soil compaction prevents moisture penetration reduces fertilizer and chemical uptake and hinders plant root growth In some cases yield losses can run as high as 30 due to compaction This new electronic soil compac tion m...

Страница 4: ...ion depths have been determined a much more effective tillage system can be employed The unit works as an efficient diagnostic tool as well as a management unit after the tillage operation has been performed One can determine how deep the soil has actually been worked and if the tillage operation has solved the compaction problems 25 When you are doing comparative analysis of the readings in a fie...

Страница 5: ...ld of view ex tends about 30o from the shaft The sensor meas ures distance by emit ting a sound wave and measuring the travel time of the first reflected wave Usually this first reflection will come from the ground just ad jacent to the shaft However if an object with a surface oriented toward the sensor is within the field of view an erroneous depth measurement may be made This can lead to depth ...

Страница 6: ...Logger 75 Full No GPS Found Sample meter power up screens with datalogger enabled left screen indicates GPS signal was found 0010 PSI 00 IN Logger OFF If when powered up the screen indicates the data logger is not activated see above screen no data will be stored by the meter The data logger can be activated through the SC 900 software see Meter Settings pg 14 23 Appendix 1 Time zone corrections T...

Страница 7: ...ration cont Meter operation cont The START button readies the meter and datalogger to take a profile meas urement It must be pressed before every attempted profile measurement START After measuring a profile pressing the REVIEW button allows the user to see the cone index value at each depth As the button is pressed the display screen will scroll sequen tially through the entire profile REVIEW Imm...

Страница 8: ...ce ca ble This cable connects to the 9 pin serial port of your computer and to the meter s computer port The meter s configuration can be modified by clicking on the Meter Settings button see Meter Settings p 14 The Com Port Meter Type Download Clear Memory and Me ter Settings buttons are explained in the Main Toolbar section p 12 SC 900 data port 21 Meter dimensions The following are the dimensio...

Страница 9: ...ent 9 T h e c o m p u t e r Communications Port to which the PC 3 5 serial cable is connected can be identified by using a paper clip 1 Disconnect the serial cable from the meter 2 To bring up the Port Selection screen click on the Com Port Button select the com port to be tested and click the Port Test button Click the Test Port Now button If the message Connection OK is displayed another device ...

Страница 10: ...al check that it has the following settings Data bits 8 Stop bits 1 Baud rate 4800 bps Parity None Timing 1 second GGA data string Spectrum GPS DGPS Cable SC900 Meter GPS Unit GPS computer interface cable Connecting to a GPS Unit 19 Warning Messages Cont This error message indi cates that the meter missed a depth reading This can be caused by a momentary but rapid insertion of the probe into the s...

Страница 11: ... cannot compute the cone index You may see this message more often when taking readings in areas with non compacted zones beneath strongly compacted zones Start Over Insertion too fast 11 Calibration Calibration Calibration Calibration Battery Change Battery Change Battery Change Battery Change Calibration The reading of the meter s load cell can be recalibrated with the following procedure Note D...

Страница 12: ...nd evenly so that there is not any side stress on the shaft If an error message is generated see Warning Messages pg 18 start over by removing probe and pushing the Start button 5 Remove probe gently If the profile was success fully measured the following message will be dis played 6 You can review compaction measurements increments by pushing the Review button You can also delete the profile by p...

Страница 13: ...en soil surface you may need to use a target to ensure reliable performance see Depth Sensor p 5 It is advisable to smooth the soil surface as much as possible before inserting the probe As the probe is inserted into the soil the 0010 PSI 00 IN N 1 Profile Information Screen 13 Download To download data from the internal data logger turn the meter off and connect the gray serial cable to the RS 23...

Страница 14: ...on If a signal is found position data will be stored along with the soil compaction data If no GPS signal is available when the logger is turned on the logger will no longer look for one when measuring and recording soil moisture data If the second box is 15 checked the logger will store the GPS value only if it has been differentially corrected If the differential correction is not found only the...

Отзывы: