Your AutoDialer is now ready for use.  It is strongly recom-
mended that you perform a test to ensure functionality.  Test 
instructions are provided in the AutoDialer Manual. 

 

Activation Hookup with Other Sensors or Control Sys-
tems

 

 
Important Notes: 

 
The power/alarm cable provided with the AutoDialer has been 
tested with up to a 400ft. extension using 24-gauge wire.  It is 
estimated this wire can be extended up to an 750ohm imped-
ance.  You can test impedance using a volt meter with imped-
ance capability.  If in doubt, test your application first! 
 
When using rainfall as an alarm actuation in SpecWare, be 
aware that the relay will not activate/deactivate unless the 
amount of rain that falls during the logging interval exceeds 
the set point. 
 
The set point seen in the SpecWare launch options screen may 
change slightly to ensure functionality with the selected pa-
rameters. 
 
 

Please refer to the AutoDialer Manual or 3rd party manufac-
turer instructions.  Wiring instructions are provided for nor-
mally open, normally closed, or voltage level activation 
setup.   