1. Lay the lamp housing on a flat, level surface.
2. Remove the 2 screws that secure the endcap opposite the
power cordset. For lamps without a filter, proceed to Step #4.
3. Slide the filter assembly off the lamp housing’s open end, and set
aside.
4. Grasp the tube by the metal bases located at each end.
Applying even pressure, gently rotate the tube a quarter of a
turn until it loosens. The tube may now be easily removed from its
sockets.
5. Install the new tube by reversing the above procedure.
6. Where applicable, replace the filter assembly by reversing the
procedure in Step #3.
7. Replace the endcap removed in Step #2 and test the function
of the new tube.
8. If installation of the new tube fails to yield correct lamp opera-
tion, check to make sure the two pins at each end of the tube
are properly connected with the two copper contacts in each
socket.
NOTE: The tubes in the EA-240 are wired in series. Therefore, if one
tube fails the other tube will not light, even though it may be in
good operating condition.
FILTER REPLACEMENT
If it is necessary to replace the filter assembly, follow Step #3 in the
‘‘TUBE REPLACEMENT’’ section to remove the old filter. Install the
new filter by reversing this procedure.
4
5
LAMP WITH
FILTER ASSEMBLY
Endcap
On/Off Power Switch(es)
(top of unit)
Filter Assembly
Power Cord
ENVIRONMENTAL CONDITIONS
The E-Series UV lamps are designed to be safe under the following
conditions:
•Indoor use;
•Altitude up to 2,000 m (6,562 ft.);
•Temperature 5°C to 40°C (41°F to 104°F);
•Maximum relative humidity 80% for temperatures up to 31°C
(88°F)
decreasing linearly to 50% relative humidity at 40°C (104°F);
•Mains supply voltage fluctuations not to exceed ±10% of the
nominal voltage;
•Installation Category II;
•Pollution Degree 2.
WARRANTY
The warranty policy for the E-Series lamps is provided on the
Certificate of Limited Warranty enclosed separately with each unit.
NOTE: For assistance of any kind, including help with a lamp under
warranty, contact the Customer Service Department at
Spectronics Corporation. In the U.S. and Canada, call toll-free
1-800-274-8888. Give full details of the difficulty and include the
model and serial numbers of the unit and date of purchase.
If return of the lamp to the factory is deemed necessary, shipping
instructions will be provided. If an estimate of charges for non-
warranty work or other service work is required, a quote will be
furnished upon evaluation of the unit. Out-of-warranty service work
will not be performed without customer approval.
SHIPPING
Carefully pack the lamp in the original shipping container and
packing materials. Insure it for full value and ship it prepaid to the
factory.