7
6
UV-A sensor
Visible sensor
Dual-sensor detector
(XDS-1000A)
Meter readout unit
(XR-1000)
•
The sealed membrane keypad on the readout meter is equipped with five pressure-
sensitive buttons that provide multiple functions and easy operation. The ON/OFF
button turns the meter on and off. The four side-by-side buttons provide intuitive
customer interactive functions.
•
The instrument will automatically turn itself off to preserve the battery life if no buttons
are pressed for 60 minutes.
Running man icon––
Meter in running operation
Bulb A icon––
Absolute readings displayed
Battery power bars
Open hand icon––
Holds data-constant display
Bulb Z icon––
Zero readings with ambient
light subtracted
Hourglass icon––
Integration of light levels––
continuous count
Mountain peak icon––
Peak value displayed
Snowflake icon––
Freezes data during integration
Circular arrows icon––
Integration in progress
4.2 DISPLAY AND ICONS
The AccuMAX meter uses a liquid crystal display (LCD) with a combination of easy-to-read icons and alphanumeric
characters to denote readings, units, operating mode and meter status in a four-line format. UV and VIS level
readings are automatically adjusted for the magnitude of the measurement (autoranging) and then displayed with the
appropriate 4-digit resolution units. For example: 12,000 µW/cm
2
would be displayed as 12.00 mW/cm
2
. This enables
the readout unit to accept and respond to a wide range of light levels. Icons provide a quick glance at data type and
the available functions at every operational screen.
4. CONTROLS
4.1 BUTTONS
Dual-wavelength
UV-A/visible sensor
detector
Water-Resistant
Adapter
LCD window with numbers,
units, modes and system status.
Displays multiplexed data from
UV-A/visible output.
3.3 ATTACHING WATER-RESISTANT ADAPTER (REMOTE CONNECTION)
The AccuMAX XRP-3000A kit includes a custom 5 ft (1.5 m) USB cable (XCB-100) with a water-resistant sensor
adapter. The male USB connector port attaches to the top of the AccuMAX readout unit and the female port inside
the adapter attaches to the bottom of the sensor detector.
With sensors facing upward, align the arrows on the adapter to the detector. Gently expose the USB port inside the
adapter by flipping the rubber edge outward, then connect it to the USB port on the sensor detector.
Note
: To ensure water resistance, make sure the two rows of ridges on the adapter are firmly in place within the
grooves of the detector.
3. INSTALLATION
3.1 CONTROLS, CONNECTORS AND INDICATORS
The AccuMAX dual-sensor detector (XDS-1000A) may be connected to the meter readout unit (XR-1000) by
direct
connection
or by
remote connection
via USB connector (see below). Either of these options allows a complete
transfer of information in digital form.
3.2 DIRECT CONNECTION
Attach detector by lining up USB connections and pushing together.