OPTIONAL ATTACHMENTS
10
1. OPACITY METER
This provides a rapid means of measuring the opacity of a
sample. Approximately 30% opacity will set the lower limit of sensitivity to 5 µm
instead of the standard 1 µm. Dilution down to 10% opacity is needed to
measure particles down to 1µm (ISO-4406) or 4µm (ISO-11171).
Installation:
1. Slide Opacity Meter “Head” in front of the receptor lens with the black
target. The circular sensor aperture should be facing the front of the
instrument, and the red laser beam should hit it fully.
2. With the dial knob, set the indicator on the Opacity Meter to “CAL”.
3. Place the beaker or bottle containing the liquid to be tested into the
“V” Block.
4. Note where the indicator on the Opacity Meter dial points. This is the
opacity of the sample.