Installation
OXY5500 Operator’s Manual
2
–
5
Protective chassis and ground connections
Before connecting any electrical signal or power, the protective and chassis
grounds must be connected. Requirements for the protective and chassis
grounds are as follows:
•
The protective and chassis grounds must be of equal or greater size
than any other current-carrying conductors, including the heater
located in the sample conditioning system.
•
The protective and chassis grounds must remain connected until all
other wiring is removed.
•
If the protective and chassis ground is insulated, it must use the
green/yellow color.
Refer to Figure 1–1 and Figure 1–2 for the protective and chassis ground
locations.
To connect electrical power to the analyzer:
1.
Open the OXY5500 analyzer electronics enclosure door. Take care
not to disturb the electrical assembly inside.
Careful consideration should be taken when grounding. Properly
ground the unit by connecting the main ground lead to the
protecting grounding stud labeled with the ground symbol .
Connect the chassis ground stud to plant grounding using 6mm
2
or 10-gauge wire.
Do not exceed the 36 VDC power rating or electronics will be
damaged.
Interconnection of the analyzer enclosure shall be accomplished
using wiring methods approved for Class 1, Division 2 hazardous
locations as per the Canadian Electrical Code (CEC) Appendix J
and the National Electric Code (NEC) Article 501. The installer is
responsible for complying with all local installation codes
or in
accordance to ATEX/IECEx standards IEC/EN 60079-0, IEC/EN IEC/EN
60079-15 and IEC/EN 60079-14
.
Hazardous voltage and risk of electric shock
.
Failure to
properly ground the analyzer may create a high-voltage shock
hazard.
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...