System Set-Up

Fig. D: Flow Restrictor Tables

(For 4:1 Concentrate to Product Ratio)

FR-25 & FR-40

ml./min.

gpd

in.

cm.

110

42

1

2.5

84

32

2

5.1

75

29

3

7.6

69

26

4

10.2

63

24

5

12.7

60

23

6

15.2

55

21

7

17.8

51

19

8

20.3

47

18

9

22.9

45

17

10

25.4

44

17

11

27.9

42

16

12

30.5

40

15

13

33.0

39

15

14

35.6

38

14

15

38.1

37

14

16

40.6

PRODUCT RATE

CUT TO LENGTH

FR-90

ml./min.

gpd

in.

cm.

269

102

1

2.5

233

88

2

5.1

213

81

3

7.6

198

75

4

10.2

183

69

5

12.7

175

67

6

15.2

164

62

7

17.8

154

58

8

20.3

148

56

9

22.9

141

54

10

25.4

136

52

11

27.9

133

50

12

30.5

129

49

13

33.0

128

48

14

35.6

124

47

15

38.1

124

47

16

40.6

PRODUCT RATE

CUT TO LENGTH

FR-60

ml./min.

gpd

in.

cm.

158

60

1

2.5

130

49

2

5.1

123

47

3

7.6

110

42

4

10.2

99

38

5

12.7

94

36

6

15.2

93

35

7

17.8

88

33

8

20.3

84

32

9

22.9

79

30

10

25.4

76

29

11

27.9

74

28

12

30.5

71

27

13

33.0

68

26

14

35.6

66

25

15

38.1

66

25

16

40.6

PRODUCT RATE

CUT TO LENGTH

FR-180

ml./min.

gpd

in.

cm.

490

186

1

2.5

460

175

2

5.1

430

163

3

7.6

400

152

4

10.2

379

144

5

12.7

356

135

6

15.2

344

131

7

17.8

326

124

8

20.3

311

118

9

22.9

300

114

10

25.4

289

110

11

27.9

281

107

12

30.5

270

103

13

33.0

263

100

14

35.6

259

98

15

38.1

256

97

16

40.6

PRODUCT RATE

CUT TO LENGTH

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

15

RO Membrane Replacement

1.  Turn off the water supply to the RO system. Place the system where

 

the membrane housing is easily accessible.

2.  Remove the black tubing from the membrane feed push-fitting 

 

by depressing the collar on the fitting with your thumb and pulling 

 

the tubing from the push-fitting (Fig. E).

3.  Lift the membrane housing from the retention clips.
4.  Unscrew the membrane housing lid. This may require two people.
5.  Use a pair of pliers to grasp the membrane stem and pull the 

 

membrane from the housing (Fig. F).

6.  Remove the black housing O-ring (Fig. F). Wash the empty housing 

 

with soapy water. Rinse thoroughly with hot, clean water.

7.  Insert new membrane into the housing, with the double

 

O-ring end first (Fig.G). The tube must fit into the recess at the 

 

bottom of the membrane housing. When the membrane is aligned 

 

with the hole, firmly push the membrane into the hole until it 

 bottoms 

out. 

8.  Place the black housing O-ring on the housing rim and carefully 

 

screw the lid back on to the base.

9.  Reconnect the black tubing to the membrane feed push-fitting. 
Note: If you have a dual-membrane system, perform steps 2 thru 9 on

 

the second membrane now.

10.  Disconnect the yellow concentrate tubing (Fig. B) from the 

 

membrane housing and remove the flow restrictor (Fig. C) 

Fig. E: Membrane Housing

blue product 
water tubing

black feed tubing

membrane feed 
push-fitting

membrane housing lid

yellow concentrate tubing

Diagnostics & 

System Maintenance

Содержание MaxPure MPDI RO/DI System

Страница 1: ...performance of membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters listed on page 3 of this manual Most municipal water supplies meet these requirements Terms and Conditions of Sale 1 To obtain service under this warranty the defective system or components m...

Страница 2: ...echanical photocopying recording or otherwise without the prior written permission of SpectraPure Inc SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 23 Replacement Parts Model Replacement Part SF MT 1 0 10 1 micron MicroTec Sediment Filter CF 10 10 10 micron Carbon Block Pre Filter SF MT 0 5 10 0 5...

Страница 3: ...s Sediment Pre Filter 1 micron MicroTec sediment pre filter Carbon Filter 10 micron carbon block pre filter RO Membrane Type Thin Film Composite TFC DI Cartridges SilicaBuster Color Indicating DI Cartridge Rejection Rate Greater than 98 average Input Water Pressure 60 psi 4 15 bar line pressure Input Water Temp 77 F 25 C Recovery Rate 20 i e 20 of the water will be collected as pure water Nominal ...

Страница 4: ...oon as the cartridge becomes exhausted Operating the Push Fittings The push fittings used on SpectraPure products can be easily serviced To remove the tubing from its push fitting a Firmly depress and hold the push fitting collar down with your thumbnail b While the push fitting collar is depressed pull the tubing straight out of the push fitting Once the tubing is removed release the collar Syste...

Страница 5: ...e Membrane Test Kit RO product water Y greater than Yes Membrane OK for Alkalinity 90 Hardness Measure feed water X Is Rejection No Replace Membrane Test Kit RO product water Y greater than Yes Membrane OK for Hardness 90 Caution This test is not to be used on softened water sources Membrane Testing SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectra...

Страница 6: ...he RO membrane SpectraPure Inc assumes no responsibility for water damage due to leaks It is the user s responsibility to determine that the system is leak free SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 19 Membrane Output Calculation Example What is the expected GPD from a 75 GPD System at 40 ...

Страница 7: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 7 System Set Up Checking the Concentrate to Purified Water Ratio This procedure will assure you of maximum life and reliability of your SpectraPure System Failure to perform this procedure can permanently damage the membranes and will void the pro rated ...

Страница 8: ...rt into the end of the yellow tubing by carefully pressing on the insert with your thumbnail Care should be taken not to crush or other wise damage the end of the capillary tubing protruding from the end of the insert SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 17 Deionization Cartridge Diagnost...

Страница 9: ...duct water tube double O ring end of membrane product water stem tube Diagnostics System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 9 System Set Up Fig B Reverse Osmosis Assembly Top Rear View membrane feed push fitting black tubing to membrane feed RO membrane yellow concentrate tu...

Страница 10: ... Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 RO Membrane Replacement 1 Turn off the water supply to the RO system Place the system where the membrane housing is easily accessible 2 Remove the black tubing from the membrane feed push fitting by depressing the collar on the fitting with your thumb and pulling the tubi...

Страница 11: ...Procedure 1 Turn on the system let it operate long enough to fill the empty DI housing then let it run for an additional 20 minutes 2 See the section on Testing the Quality of the Membrane page 19 Perform the test using your chosen method 3 Turn off the system drain the DI housing re install the DI cartridge and turn on the system Note All water sources are different and are subject to changes in ...

Страница 12: ...ectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 Carbon Block Filter Diagnostic Replace the Carbon Block Filter at least every 6 months OR when chlorine breakthrough greater than 0 1 ppm occurs in the yellow con centrate line To test for chlorine breakthrough collect a 10 ml sample of the concentrat...

Отзывы: