Procedure:
1. When the reading on the right-hand TDS meter (set to “IN”) displays 75%
of the reading on the left-hand TDS meter (set to “OUT”), it is time to
replace
the
M
AX
C
AP
DI
™
cartridge
.
2. Follow the directions on page 5 to rinse up the
M
AX
C
AP
DI
™
cartridge
.
3. Make sure the DI cartridge is installed in the correct direction as marked
on the cartridge shell and be sure that the top seal is securely attached to
the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
Procedure:
1. When the reading on the right-hand TDS meter (set to “OUT”) displays
“001”, it is time to replace the SilicaBuster
™
DI cartridge.
2. Follow the directions on page 5 to rinse up the SilicaBuster
™
DI cartridge.
3. Make sure the DI cartridge is installed in the correct direction as marked
on the cartridge shell and be sure that the top seal is securely attached to
the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
For service or repair of these monitors, please send to:
HM
DIGITAL,
INC
5819
Uplander
Way
Culver City, CA 90230
DM-1 METER Specifications:
Range
0-1999
PPM
Resolution
1PPM (1-999 PPM)
Accuracy
2%
Probe
¼’’
Power Source
(2) 1.5V button batteries
Battery life
Approx 1000 hours
MAXCAP DI MAINTENANCE AND REPLACEMENT:
SILICA-BUSTER MAINTENANCE AND REPLACEMENT: