SANITIZING THE RO SYSTEM (once a year)
Sanitizing is recommended at least once every year or if water smells (or tastes) bad even after a post carbon filter
replacement. A convenient time for sanitization is during a filter change-out. IT IS IMPORTANT THAT YOU HAVE CLEAN
HANDS WHILE HANDLING INNER PARTS OF THE SYSTEM.
1. Be sure water supply to the RO/DI system is turned off, and the dispenser faucet is open. This will completely drain the
pressure tank.
2. Next, remove the 1/4” black tubing from the feed port of the membrane and lift membrane housing from clips.
3. Remove the membrane housing cap by unscrewing it counterclockwise. Grasp the membrane stem with a pair of
pliers and pull the membrane from the housing.
4. Screw the cap back on to the membrane housing making sure that the o-rings are in place.
5. Reconnect the 1/4” black tubing to the feed port of the membrane.
6. Remove sediment, carbon block, deionization cartridge and in-line post carbon filter.
7. Put 1.5 - 2.0 oz. of household bleach in the left filter housing and fill it half-way with tap water. Now, put all filter
housings back on the bracket, keeping the one with the bleach on the left side.
8. Next, close the dispenser faucet (put lever in DOWN position).
9. Open feed supply valve 1/4 turn from closed position.
10. Allow 15 minutes for the bleach solution to flow through the system. Then, open dispenser faucet and keep the lever
in the UP position till some bleach solution is dispensed through the faucet. IMMEDIATELY close the faucet as soon as
the bleach solution is dispensed. This will sterilize the faucet and the line going to the faucet.
11. Let the system sit for 2 hours.
12. Open the dispenser faucet to drain bleach solution from the pressure tank as completely as possible and then close
the faucet.
13. Open the feed supply valve fully.
14. Allow the tank to fill until pressure gauge reaches 40 psi. Then open the dispenser faucet and flush system until all
bleach solution has been dispensed from the system.
15. Close the feed supply valve.
16. Remove filter housings and membrane housings from the system and then rinse them with tap water and drain them
completely.
17. Put new replacement filters (sediment filter, carbon block filter & in-line post carbon filter) into the filter housings and
put the membrane back into the membrane housing. You can use the same deionization cartridge if DI cartridge has
not exhausted.
Open the feed supply valve and check for leaks. This completes the procedure.
RECHARGING THE BLADDER TANK
If the storage capacity of the tank is diminished significantly it is likely that the tank has lost its air charge. Recharging the
bladder tank will restore its capacity.
Be sure water supply to the RO/DI system is turned off, and the dispenser faucet is open. This will completely drain the
pressure tank.
1. Put the dispenser faucet in the “open” position (lever in up position). Leave the faucet in the open position until the
procedure is completed.
2. Drain as much water as possible from the tank.
3. Hook up an air pump to the “schrader” (tire) valve on the pressure tank and start pumping air into the tank.
4. Expel all the water from the tank.
5. Continue pumping air into the tank until the pressure reads NO MORE than 6 PSI.
6. This completes the procedure. Turn system back on.
Note: Should this procedure fail to restore the capacity of the tank, it is very likely that the bladder in the tank is ruptured
and the tank needs replacement.
MAINTENANCE PROCEDURES