Internal oxygen concentration result drops below 0.1%, 
but there is considerable variation in “Self check” results. 
Once the result is very low, around 0.01–0.05%, the op-
erator might see some variation in the “Self check” read-
ings so that some readings might be even > 0.1%. This is 

  Settings page after  

  Self check is performed

normal, as the concentration is so low, that sometimes 
the measurement noise can cause variation. Important 
thing is that most of the “Self check” measurements are 
< 0.1%, as required.

5.8 Measurement problems

If unexpected inconsistent readings are received, please 
check the following things:

 

Î

Laser temperature

 

Î

Internal oxygen concentration

 

Î

Vacuum suction holds the measuring head tightly 
against the insulating glass to be measured

 

Î

Glass and spacer thicknesses are approximately 
according to the correct expected thicknesses

 

Î

Try to measure close to the spacer.

 

Î

Do several measurements on the same IGU and 
compare the results. If most of the measurements 
are consistent but there are exceptional results, 
repeat the measurement. The exceptional results 
might be measurement errors.

 

Î

If there is a special reflective coating (e.g., Low-E), try 
measuring from the non-coated side to get a better 
signal to the sensor and more accurate results. 

 

Î

If the vacuum does not hold the measuring head 
tightly against the glass, make sure that the IG 
surface is not too bent to avoid proper contact. 
You can try to help the vacuum by pressing the 
measuring head against the glass firmly once the 
measurement has started.

 

Î

If glass or spacer thicknesses are shown incorrectly, 
the measurement is taken from the incorrect 
location and the results are not correct. The device 
operator can try to help the analysis by providing 
the pre-set values to the drop-down menus for glass 
and spacer thicknesses on the Settings Page. When 
the operator provides this information, the software 
can locate the correct measurement locations more 
easily.

If the problems persist, please contact 

 

[email protected]

. Also, please be prepared to 

have the IG structure information, coating characteris-
tics available, and the measurements data file.

11

Содержание Laser Portable 2.0

Страница 1: ...ner not specified by the manufacturer the safety protection of the equipment may be compromised Î Do not replace detachable mains supply cords with inadequately rated cords Î Always make sure that you are using a safety grounded socket with a safety grounded mains supply cord to connect the mains supply Î Never flush the device with a gas flow greater than 3 litres min Chapter 3 2 Î To ensure the ...

Страница 2: ...on the device from the power switch on the main unit A blue light will light up around the power switch Windows will start and Start buttons led on the measuring device s handles will light up After Windows has started start Laser Software by Sparklike When starting the Laser software the lights at the joint of the handles should turn on NOTE When the device is turned on start the software and let...

Страница 3: ...products To measure triple glazed units slide down the button on the left hand side of the Main Page Once this is done the Triple IG indicator lights up The default settings can be found in the image representing the Main Page with the default setting 3 Pressing the Measure button will initiate the measurement During the measurement the Measuring indicator is blinking After the measurement has bee...

Страница 4: ...down box indicates the time to collect data from one cavity This can be adjust ed between 2 5 seconds The default time is 5 seconds Longer measurement time is advised for greater accu racy Enable counter and Meas count 3 refer to settings for making multiple measurements discussed below in Section 4 3 1 High limit and Low limit 4 refer to the desired result range For easy interpretation of the res...

Страница 5: ...highly accurate measurement results Let the device warm up with software running for 15 30 min Incorrect temperature adjustment might result in incorrect measurement results or unexpected functions Take the following actions to test the temper ature adjustment whenever starting up the device Choose the Test page from the upper left corner of the software Press the Laser I R button the indicator be...

Страница 6: ...nlet is next to the power switch Attach the connector for the flushing gas hose to the main unit Join the flushing gas hose 6x4mm Î Apply dry and pure 99 9 flushing gas flow of 1 3 l min maximum 3 l min to the hose Î Press the Gas fill button on the software light on the button will turn on This will open the internal valves and guide the flushing gas flow through the measuring head During the fil...

Страница 7: ... the Main Page turns red when measuring This reminds the user that the measurement settings are different from the default values To return all the values back to default values press the Reset button on the Main Page or change the values back manually When pressing the Reset button it takes a few seconds to return all the values back to default 4 3 2 Editing the default settings You can edit the ...

Страница 8: ...c C Temp Results 5 TROUBLESHOOTING 5 1 Restarting the program When starting the system if the indicator leds on the handles of the measuring head do not light up Chapter 2 restart the program by closing the window from the upper right corner Start the program by double clicking the Laser software icon on your desktop If the Windows freezes and you are unable to shut down the device press the power...

Страница 9: ...ues are correct the device is firmly against the glass and the device is not tilted In these cases the device window may need to be cleaned If that does not help shut down the Win dows and wait 20 seconds before restarting the device If the above mentioned actions do not work error message is caused by the lack of the laser signal on the detector This is due to insufficient amount of laser light p...

Страница 10: ...ser that works by meas uring the oxygen inside the glazing unit To ensure cor rect readings oxygen needs to be removed from inside the measuring head by replacing it with argon or other flushing gas such as xenon or nitrogen Internal oxygen concentration is tested from the Settings Page on the software by pressing the Self check button When testing the internal oxygen level or making the argon fil...

Страница 11: ...easurements are consistent but there are exceptional results repeat the measurement The exceptional results might be measurement errors Î If there is a special reflective coating e g Low E try measuring from the non coated side to get a better signal to the sensor and more accurate results Î If the vacuum does not hold the measuring head tightly against the glass make sure that the IG surface is n...

Страница 12: ...asuring device and the device must be packed appropriately to ensure safe shipping Go to sparklike com guidance lp to get instructions how to pack the device for transpor tation The window of the measurement head is sensitive to scratches If the window of the measurement head requires cleaning use materials and solvents made for delicate optical instruments Ethernet connector location in the conne...

Страница 13: ...indoor use only Humidity 20 80 Altitude up to 2000 m Thickness measurement accuracy 50 μm Repeatability of the gas measurements 2 Measurement time 18 28 s depends on the settings and IG structure Maximum IG thickness 51 mm from surface 1 to surface 3 with DGU from surface 1 to 5 with TGU Minimum panel thickness 2 mm Connectors 4x circular multipin connector RJ 45 2x USB 2 0 VAC Power C14 Display 1...

Страница 14: ...SPARKLIKE OY www sparklike com Hermannin rantatie 12 A 21 00580 Helsinki Finland 05 2021 Scan the QR code for further instructions and support sparklike com guidance lp ...

Отзывы: