www.spacelabshealthcare.com
4 - 3 3
9 1 3 9 0
Q U B E
M
A I N T E N A N C E
Figure 4-30 Latch Docking Station
7
Remove the six (6) connector standoff screws from the back of
the docking station with a nut driver. Refer to
.
Figure 4-31 Docking Station Connectors
8
Turn the docking station back over. Carefully slide out the
docking station PCBA. Do not damage the connectors.
9
Install a new docking station PCBA in the reverse order.
10
Perform functional tests in accordance with the monitor
functional tests. Refer to
Replace the Docking Station Latch
1
Disconnect all cables from the docking station.
2
Turn the docking station over. Remove the six screws from the
docking station cover plate. Refer to
3
Remove the docking station cover plate. Set it aside.
4
Remove the connection J70 (LED wiring assembly) from the
docking station PCBA. Refer to
.
Содержание qube 91390
Страница 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Страница 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Страница 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Страница 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Страница 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Страница 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Страница 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...