The embedded front and rear alarm light LEDs illuminate
briefly. Verify that all LEDs are functional.
c
The
Diagnostic
menu displays. (If errors are noted during
power ON, contact your Spacelabs Healthcare Field Service
Engineer.)
d
If batteries are installed, the fan begins operating within
60 seconds after turning ON the monitor. The fan turns OFF
unless the internal temperature requires more cooling by the
fan.
e
Monitor keys display along the right side of the screen.
4
Insert the ECG module without a patient cable connected. Verify
that the ECG parameter key appears on the screen with
???
and
the message
LEADS OFF
.
5
Connect a patient simulator to the ECG input with a patient cable.
Set the simulator to a known rate.
6
Verify that the ECG count and the lead being monitored are
displayed to the right of the ECG parameter key.
7
Verify that the ECG waveform appears on the screen.
8
Connect a patient simulator to the invasive pressure inputs.
9
Zero the pressures and verify that the numerics and waveforms
are accurate.
Optional Recorder Assembly for Options A and B
If the monitor is configured with the optional recorder assembly,
verify that the monitor initiates a recording. To verify that the
monitor initiates a recording, touch the RECORD key then touch one
of the flashing parameter keys.
Note:
The printed waveform should not have defects such as gaps,
extra lines, etc. If there are defects, notify a Spacelabs Healthcare
Field Service Engineer for servicing, or replace the recorder.
External Alarm Relay Output for Option A
If the monitor is configured to use the alarm relay output, to
interface to an external alarm device:
1
Plug the cable into the external alarm connector.
2
Initiate a high-priority alarm.
3
Verify that the external alarm responds appropriately.
4
If an external alarm light capable of displaying multiple colors is
attached, turn the high-priority alarm OFF. Generate a medium-
priority alarm.
5
Verify that the external alarm responds appropriately.
6
Repeat steps 4 to 5 for a low-priority alarm.
Содержание qube 91390
Страница 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Страница 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Страница 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Страница 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Страница 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Страница 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Страница 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...