Pathfinder SL v1.9 Operations Manual
18-10
18.13 QT Rejection Criteria
Noise will perturb the QT detection algorithm and so it is important to exclude (automatically
values that exceptionally long or short using the QT Value Range.
QT analysis becomes unreliable at high heart rates due to the P wave encroaching on the end
of the T wave. The Max HR should be set at 150 or less for adults but may be increased for
neonates, for example.
18.14 User Profiles
Once you have determined and set your different advanced options, you may save them in a
user profile.
Содержание 98000
Страница 4: ...Pathfinder SL v1 9 Operations Manual iv This page is intentionally blank...
Страница 16: ...Pathfinder SL v1 9 Operations Manual 1 6 This page is intentionally blank...
Страница 34: ...Pathfinder SL v1 9 Operations Manual 4 8 This page is intentionally blank...
Страница 42: ...Pathfinder SL v1 9 Operations Manual 5 8 This page is intentionally blank...
Страница 46: ...Pathfinder SL v1 9 Operations Manual 6 4 This page is intentionally blank...
Страница 50: ...Pathfinder SL v1 9 Operations Manual 7 4 This page is intentionally blank...
Страница 56: ...Pathfinder SL v1 9 Operations Manual 8 6 This page is intentionally blank...
Страница 62: ...Pathfinder SL v1 9 Operations Manual 9 6 This page is intentionally blank...
Страница 72: ...Pathfinder SL v1 9 Operations Manual 12 2 This page is intentionally blank...
Страница 78: ...Pathfinder SL v1 9 Operations Manual 14 4 This page is intentionally blank...
Страница 96: ...Pathfinder SL v1 9 Operations Manual 17 4 This page is intentionally blank...
Страница 114: ...Pathfinder SL v1 9 Operations Manual 19 6 This page is intentionally blank...
Страница 120: ...Pathfinder SL v1 9 Operations Manual 20 6 This page is intentionally blank...
Страница 121: ......