Pathfinder SL v1.9 Operations Manual
18-4
To correct measurements, let the mouse cursor hover over the vertical line until the double
headed arrow appears, then drag the line to the new position.
It is possible to apply changes to the individual beat or to all subsequent beats by using the Till
Stop function.
Periods may be excluded from analysis using the Skip function.
6. The Trend display indicates
where a value has been
excluded from analysis. (The
colour of the vertical line
depends upon the exclusion
reason). It is also possible
to see the relationship
between the QT Peak and
QT end values.
Clicking on the trend will display the corresponding ECG complex. If the measurement is
valid, these will be reported in the boxes immediately below the trend. If, as above, the
measurement has been excluded, then a zero value will result for QT and QTc.
Beats may be validated or invalidated and the measurement cursors repositioned when
analysis is halted, as previously described.
7. At the end of analysis, press the End Analysis
button and at the next screen, click on OK
to accept the results of analysis.
8. When QT analysis has finished, the QT reports are listed on the left pane of the report
window. The report formats are the same regardless of the type of analysis performed.
The navigation pane on the left hand side of the screen lists all of
the reports available for viewing.
To view a particular report simply click on the report name.
Alternatively, click on the blue arrow
→
on the toolbar to show
the next type of report in the list.
The Histogram, Trend and Strips display all have editing tools for correction of analysis. It is not
possible to edit from the other screens.
Note
All the report screens have these icons in the toolbar. From let to right, these allow you to:
• Show or hide the navigation pane
• Access the Options menu
• Display patient information
Note
that Tamp = T wave amplitude, and DI is defined as DI = RR-QT.
DI stands for Diasystolic Interval.
Содержание 98000
Страница 4: ...Pathfinder SL v1 9 Operations Manual iv This page is intentionally blank...
Страница 16: ...Pathfinder SL v1 9 Operations Manual 1 6 This page is intentionally blank...
Страница 34: ...Pathfinder SL v1 9 Operations Manual 4 8 This page is intentionally blank...
Страница 42: ...Pathfinder SL v1 9 Operations Manual 5 8 This page is intentionally blank...
Страница 46: ...Pathfinder SL v1 9 Operations Manual 6 4 This page is intentionally blank...
Страница 50: ...Pathfinder SL v1 9 Operations Manual 7 4 This page is intentionally blank...
Страница 56: ...Pathfinder SL v1 9 Operations Manual 8 6 This page is intentionally blank...
Страница 62: ...Pathfinder SL v1 9 Operations Manual 9 6 This page is intentionally blank...
Страница 72: ...Pathfinder SL v1 9 Operations Manual 12 2 This page is intentionally blank...
Страница 78: ...Pathfinder SL v1 9 Operations Manual 14 4 This page is intentionally blank...
Страница 96: ...Pathfinder SL v1 9 Operations Manual 17 4 This page is intentionally blank...
Страница 114: ...Pathfinder SL v1 9 Operations Manual 19 6 This page is intentionally blank...
Страница 120: ...Pathfinder SL v1 9 Operations Manual 20 6 This page is intentionally blank...
Страница 121: ......