Operation
MAX
MIN
n m VA F M k
AUTO HOLD MAX MIN REL LOZ
0
10
20
30
40
1. Set the rotary function switch to
the position.
2. Press the MODE button to select
frequency or % duty cycle. The “Hz”
or “%” symbol will appear on the
LCD display.
3. Insert the black test lead into the
COM
input jack and the red test
lead into the
V
input jack.
4. Touch the test lead probes to the circuit under test.
5. Read the frequency or % duty cycle on the LCD display.
Frequency and % Duty Cycle Measurements
Observe all safety precautions when
working on live voltages. Do not connect to circuits that
exceed 250V when the meter is set to Frequency or Duty Cycle.
WARNING:
1. Set the rotary function switch
to the position.
2. Insert the black test lead into the
COM
input jack and the red test
lead into the
10A
input jack.
3. Remove power from the circuit
under test, then open up the
circuit at the point where you wish
to measure current.
4. Touch the test lead probes in series with
the circuit being measured.
5. Apply power to the circuit.
6. Read the current on the LCD display.
AC Current Measurements
Observe all safety precautions when
working on live circuits. Do not measure current on circuits
that exceed 600V. Measurements in the 10A range should
be limited to 30 seconds maximum every 15 minutes.
WARNING:
AC
MAX
MIN
n m VA F M k
AUTO HOLD MAX MIN REL LOZ
0
10
20
30
40
MAX
MIN
n m VA F M k
AUTO HOLD MAX MIN REL LOZ
0
10
20
30
40
MAX
MIN
n m VA F M k
AUTO HOLD MAX MIN REL LOZ
0
10
20
30
40
-
+
10
Operation cont.
1. Set the rotary function switch to
the position.
2. Insert the black test lead into the
COM
input jack and the red test
lead into the
10A
input jack.
3. Remove power from the circuit
under test, then open up the circuit
at the point where you wish to
measure current.
4. Touch the test lead probes in series
with the circuit being measured.
Touch the red probe to the positive
side of the circuit and touch the black
probe to the negative side of the circuit.
5. Apply power to the circuit.
6. Read the current on the LCD display.
DC Current Measurements
Observe all safety precautions when
working on live circuits. Do not measure current on circuits
that exceed 600V. Measurements in the 10A range should
be limited to 30 seconds maximum every 15 minutes.
WARNING: