ST2523 Series Operation Manual Ver1.2
27
5.1.7
Display Digits
SLOW, MED
FAST
Resistance range
5 digits, maximum display
number: 35,000
4 digits, maximum display
number: 3500
Voltage range
5 digits, maximum display
number: 31,000
4 digits, maximum display
number: 3100
5.2
Test Signal
5.2.1
Frequency
The test signal is a sinusoidal wave with the frequency of 1kHz±0.2Hz.
5.2.2
Maximum Input Voltage
ST2523: The maximum input voltage that the test terminal can withstand is 65V.
ST2523A: The maximum input voltage that the test terminal can withstand is 310V.
5.2.3
Maximum Display Range
Parameter
Measurement Display Range
L
0.2000nH to 500mH
R, X, Z
0.001mΩ to 3.5000kΩ
Q
0.001 to 9999.9
θ
Deg -179.99° to 179.99°
Rad -3.1416 to 3.1416
V (ST2523/A)
0.0001V to 310.000V (ST2523A)
5.2.4
Measurement Accuracy
Measurement accuracy depends on measurement stability, temperature coefficient, linearity, repeatability, and
correction interpolation error.
Checking the measurement accuracy should be taken under the following circumstances:
a
.
Warm-up time should be more than 30 minutes.
b
.
Correctly short the test cables and reset the instrument by pressing the REL button on the
front panel.
The correct shorting of the test cable is done as follows: