136 | Service Manual: FUSION EQUINE
DR®
5. Maintaining the FUSION EQUINE DR
®
X-ray
System
Performing Panel Calibration
The Sound user can perform gain calibration on the active panel. Set the gain calibration
frequency Config> Advanced Options screen. The calibration consists on 4 Successful
exposures.
Prerequisites
Before you begin this task, ensure that the panel is installed, configured, and active on the x-
ray system.
Procedure
1.
Open the
Management
Displaying the Management screen
on page
64, for instructions.