SORVALL
®
Centrifuges
Operation
3-5
2 .
Wh en t h e
OPEN
ligh t com es on , t u r n t h e d oor r elea s e k n ob t o t h e
OPEN
p os it ion , a n d
lift t h e ch a m b er d oor u p .
3 .
In s t a ll t h e r ot or on t h e d r ive s h a ft a n d clos e t h e d oor. Tu r n t h e d oor r elea s e k n ob t o
t h e
LOCKED
p os it ion (t h e d oor la t ch es will en ga ge).
CAUTION
Be sure the rotor is properly balanced and seated on the drive spindle. See
rotor instruction manual.
4 .
R T6 0 0 0 B o n ly :
S et t h e
CHAMBER TEMP
°
C
d ia l t o t h e r u n t em p er a t u r e d es ir ed .
5 .
S et t h e
BRAKE
s wit ch t o " " if b r a k in g is d es ir ed .
6 .
S et t h e
TIME
d ia l t o t h e r u n t im e d es ir ed .
7 .
S et t h e
REV/MIN x 1000
d ia l t o t h e d es ir ed r u n s p eed .
NOTE
The settings on the
REV/MIN x 1000
dial indicate actual rotor speed. Verify
the rotor speed by checking the
REV/MIN x 1000
indicator.
A zero speed switch is incorporated in the
REV/MIN x 1000
dial. This switch
prevents rotation of the rotor if the door is locked and the timer is on.
At t h e en d of t h e r u n r em ove t h e r ot or a s d es cr ib ed b elow.
WARNING
Alw a y s
check that the rotor has stopped spinning
b e fo re
opening the
chamber door. Do not open the door while the rotor is still in motion; to
do so can result in possible injury.
To remove the rotor:
1 .
Lea ve t h e
POWER
s wit ch s et t o "
I
".
2 .
If t h e
TIME
d ia l is s et t o
∞
, t u r n it off.
3 .
Wh en t h e
OPEN
ligh t com es on , t u r n t h e d oor r elea s e k n ob t o
OPEN
p os it ion a n d lift
t h e ch a m b er d oor u p .
4 .
Rem ove t h e r ot or, clos e t h e ch a m b er d oor, a n d s et t h e
POWER
s wit ch t o "
0
".
Содержание RT6000B
Страница 25: ...SORVALL Centrifuges Maintenance 4 7...