SORVALL
®
Centrifuges
Operation
3-3
Air flow t h r ou gh t h e T6 0 0 0 B cen t r ifu ge is d es ign ed t o m in im ize in cr ea s e in r ot or
t em p er a t u r e. Air en t er s t h r ou gh a n op en in g in t h e cen t er of t h e lid a n d exit s t h r ou gh t h e
exh a u s t d u ct a t t h e t op of t h e r ot or ch a m b er.
WARNING
Blocking the air flow entering and/or exiting the T6000B rotor chamber
will cause an increase in the temperature within the rotor chamber. The
temperature increase can temporarily distort the non-metal surfaces
allowing particles to exit the rotor chamber resulting in personal injury and/
or centrifuge damage should tube breakage or rotor failure occur.
Wh en r u n n in g t em p er a t u r e-s en s it ive m a t er ia l we r ecom m en d t h a t you d o a t r ia l r u n .
b. Rotor Installation, Loading and Balancing
Refer t o t h e r ot or in s t r u ct ion m a n u a l for loa d in g a n d b a la n cin g p roced u r es a s well a s
in for m a t ion r ega r d in g t h e s elect ion a n d u s e of t u b es , b ot t les , a n d a d a p t er s .
WARNING
When loading the rotor, be sure not to exceed the maximum compartment
mass of the rotor (see Table 1-1 or the individual rotor instruction manual).
If maximum compartment mass is exceeded, maximum rotor speed must be
lowered as described on page 3-6, Reducing Speed for Rotor Compartment
Loads in Excess of Design Mass. Failure to do so can cause rotor failure
which could result in personal injury and/or centrifuge damage.
CAUTION
Do not operate the centrifuge without a rotor installed and properly
balanced. Also, when installing a rotor carefully place it on the centrifuge
drive spindle. The drive can be damaged if rotors are dropped onto it.
Precooling a Rotor in the RT6000B Centrifuge
For op er a t ion a t t em p er a t u r es ot h er t h a n a m b ien t , t h e r ot or a n d r ot or ch a m b er s h ou ld
b e p r ecooled b efor e t h e r u n u s in g eit h er of t h e p r ecoolin g m et h od s d es cr ib ed .
a. Precooling a Rotor in the Centrifuge
1 .
Follow s t ep s 1 t h r ou gh 6 of t h e op er a t in g in s t r u ct ion s (p a ges 3 -4 a n d 3 -5 ).
2 .
S et t h e
REV/MIN x 1000
d ia l t o 1 5 0 0 r p m .
Th e len gt h of t im e it will t a k e t o p r ecool a r ot or will va r y d ep en d in g on t h e t em p er a t u r e
s elect ed , r ot or weigh t , a n d r ot or m a t er ia l.