21. PROGRAMMABLE PARAMETERS
21.3.3 .
Therapy parameters in fast VT / VF zone
ATP 1 program
Values
ATP program
OFF-
Burst
-Burst+Scan-Ramp-Ramp+Scan
Number of sequences
1
-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-
8
-9-10-11-12-13-14-15
Cycles added per sequence
0
-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-
80
-85-90-95
Ramp decrement (per cycle) (ms)
0
-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0
-4-8-12-16-20-30-40-50-60
Time limit
10s-20s-
30s
-1min-1.5min-2min
Minimum cycle length (ms)
95-110-125-140-155-170-190-
205
-220-235-250-265-280-
295-310
Shock program
Values
Shock 1 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Shock 2 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Number of Max. Shock (42 J)
1-2-3-
4
21.4.
REMOTE ALERTS AND WARNINGS
The device routinely performs security self-checks and technical measurements to ensure
system integrity. When system integrity is found to be at risk outside a follow-up, alerts are
stored in the device memory. When system integrity is found to be at risk during a follow-up,
the information is managed as a warning (pop-up message) to notify immediately the user.
For example, the following types of event can trigger a warning or an alert: technical
problem during a shock, pacing lead impedance or coil impedance measurements out-of-
range, battery depletion, etc. The Remote tab presents an overview of all the alerts
managed by the device.
General parameters
Values
RF communication
(1)
ON-
OFF
Remote alerts
(1)
ON
-OFF
(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.
When Alerts are programmed "On", the following System Alerts are automatically activated:
─
Battery depletion – RRT
─
Device reset
─
Excessive charge time (>25s)
SORIN – INTENSIA SonR CRT-D 184 – U150A
59