 16.  MAIN FUNCTIONS

16.8.3 .

Data transmitted

The data transmitted are identical to the data available during a standard interrogation with 
the dedicated programmer. All counters, histograms, IEGMs and diagnosis available in the 
device are transmitted containing (not exhaustive list):

programmed parameters

Information on patient and system implanted

battery status

lead status (brady leads and defibrillation coils) 

pacing counters and mean heart rate (brady)

atrial and ventricular arrhythmia counters and episodes

ventricular therapy counters

Data   are   presented   in   the   form   of   2   reports   to   the   physician:   the   first   one   contains   a 
summary of major counters, histograms, warnings and diagnosis. The second one presents 
the 3 most important IEGM episodes automatically selected based on the degree of severity 
for the patient.

16.8.4 .

User website

On the website, the physician is able to:

consult and schedule the remote follow-ups of their patient

configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail

consult, print and export patient reports 

16.8.5 .

Alert system

The   following   set   of   alert   trigger   can   be   independently   programmed   ON/OFF   by   the 
physician using the dedicated programmer and can trigger an alert transmission:

Low or high impedance (A, RV, LV)

Abnormal coil impedance (shock lead)

Low or High shock impedance

Inefficient high energy shock

All shocks programmed OFF

Shock treated VT/VF

Lack of V pacing in CRT device

Suspicion of noise on the V lead

AT/AF occurrence

Fast V rate during AT/AF

42

SORIN – INTENSIA SonR CRT-D 184 – U150A

WARNING: The use of remote monitoring does not replace regular follow-up. Therefore, 
when   using   remote   monitoring,   the   time   period   between   follow-ups   visits   may   not   be 
extended.

Содержание Intensia SonR CRT-D 184

Страница 1: ...Implant manual Implantable cardioverter defibrillator SonR CRT D model 184...

Страница 2: ...blank...

Страница 3: ......

Страница 4: ...blank...

Страница 5: ...Conformance to standards 23 10 Physician guidelines 26 10 1 Physician training 26 10 2 Directions for use 26 10 3 Maintaining device quality 26 10 4 V V Programming Recommendation 27 11 Patient infor...

Страница 6: ...ime RRT 44 17 4 Explantation 44 17 5 Defibrillator identification 45 18 Supplemental Information 46 18 1 Adverse events in the SafeR AAI DDD study 46 18 2 SafeR AAI DDD clinical study 47 19 Physical c...

Страница 7: ...ccelerometer to allow adaptation of pacing to suit the patient s activity INTENSIA SonR CRT D 184 is also equipped with the RF wireless technology which enables to remotely monitor the patients who ha...

Страница 8: ...tion for automated treatment of life threatening arrhythmias The device is also indicated for the reduction of heart failure symptoms in medically optimized NYHA Functional Class III and IV patients w...

Страница 9: ...ion digitalis intoxication drowning electrocution electrolyte imbalance hypoxia sepsis or unstable ischemic episodes who present incessant tachyarrhythmia who have an internal pacemaker whose primary...

Страница 10: ...vailable Electrical Isolation Do not permit the patient to contact grounded equipment that could produce hazardous leakage current Ensuing arrhythmia induction could result in the patient s death Disa...

Страница 11: ...2 Before procedure disable ATP and shock therapies 3 During the procedure keep the electrocautery device as far as possible from the cardiac defibrillator Set it at minimum intensity Use it briefly 4...

Страница 12: ...plant the device after this date because the battery may have reduced longevity and sterility may be affected It should be returned to Sorin If Package is damaged Do not use the device or accessories...

Страница 13: ...either a conventional atrial lead without SonR capability or a SonR atrial lead bipolar pacing sensing and SonR capability IS 1 refers to the international standard whereby leads and generators from d...

Страница 14: ...e delivery of energy to the heart Place the sutures well outside the coil of the patch lead or in the area between the coils to avoid possible coil fracture If countershock is unsuccessful using exter...

Страница 15: ...IO CRT D are applicable to INTENSIA SonR CRT D 184 The table below summarizes the adverse events observed for the CRT D system No deaths were related to the system Event of Patients of Patients of Eve...

Страница 16: ...Acceleration of arrhythmias caused by device Air embolism Bleeding Chronic nerve damage Erosion Excessive fibrotic tissue growth Extrusion Fluid accumulation Formation of hematomas or cysts Inappropri...

Страница 17: ...5 ADVERSE EVENTS Fear of premature battery depletion Fear of shocking while conscious Fear that shocking capability may be lost Imagined shocking phantom shock SORIN INTENSIA SonR CRT D 184 U150A 15...

Страница 18: ...of life for CRT D patients than for control patients System complication free rate 67 at six months 6 1 2 Methods Patients were New York Heart Association class III or IV and had one or more indicati...

Страница 19: ...est group n 91 11 84 3 90 3 50 26 3 13 41 4 28 6 18 27 67 1 57 Absolute difference between test and control groups change in QOL score over 6 months Baseline Mean SD range 6 month Mean SD range Differ...

Страница 20: ...nd OVATIO CRT D 6750 values 0 to 63 ms in steps of 8 ms The graph below shows the programmed V V settings at randomization by percentage of patients programmed to each combination of Synchronous BiV p...

Страница 21: ...remaining 174 patients 154 89 were tested per the V V optimization protocol One hundred forty nine 149 of the 154 patients who were tested per the V V optimization protocol were programmed per the rec...

Страница 22: ...tients Antiarrhythmic drug therapy If the patient is being treated with antiarrhythmic or cardiac drugs the patient should be on a maintenance drug dose rather than a loading dose at the time of ICD i...

Страница 23: ...here has been no quantitative assessment of the presence of leachables in breast milk Pediatric Patients This device has not been studied in patients younger than 18 years of age Geriatric Patients Mo...

Страница 24: ...ng points in counselling the patient about this device Persons administering CPR may experience tingling on the patient s body surface when the patient s ICD system delivers a shock Advise patients to...

Страница 25: ...ic and electrical apparatus with the basic restrictions related to human exposure to electromagnetic fields 10 MHz 300 GHz EN 301 489 1 v1 8 1 EN 301 489 27 v1 1 1 Electromagnetic compatibility and Ra...

Страница 26: ...used to communicate weather data the Meteorological Satellite or the Earth Exploration Satellite Services and must accept interference that may be caused by such stations including interference that m...

Страница 27: ...susceptible de compromettre le fonctionnement du dispositif Conform ment la r glementation d Industrie Canada le pr sent metteur radio peut fonctionner avec une antenne d un type et d un gain maximal...

Страница 28: ...mplication rates Established patient follow up Maintenance of competence 10 ICD and CRT D procedures per year 20 patients per year in follow up 1 Please consult full text of both publications for deta...

Страница 29: ...to 122 F 50 C as an electrical reset condition may occur Use Before date has expired because this can adversely affect pulse generator longevity or sterility 10 4 V V PROGRAMMING RECOMMENDATION It is...

Страница 30: ...e outer storage package Additional copies can be obtained by contacting your Sorin representative or on the Sorin s web site http www sorin com This information should be given to each patient with th...

Страница 31: ...ANTY AND REPLACEMENT POLICY Sorin warrants its defibrillators Refer to the section Warranty for additional information Please see the following labelling sections for information concerning the perfor...

Страница 32: ...hing function are available INTENSIA SonR CRT D 184 enables an adjustment of the interventricular delay and provides the possibility of adapting pacing to each ventricle The ICD is intended to resynch...

Страница 33: ...ctor head has three ports Atrial SonR port performs atrial bipolar pace sense if a conventional IS 1 lead is connected LV IS 1 port performs left ventricular bipolar pace RV DF 4 port performs right v...

Страница 34: ...ckaging contains a defibrillator and one screwdriver The non sterile items contained in the outer storage package are the implant manual the ICD Registration Form and its envelope the patient booklet...

Страница 35: ...ated before implantation Verify all shock therapies are disabled in order to avoid accidental discharge during implantation It is not advisable to program the Smoothing function before implantation si...

Страница 36: ...ramming When active case and SVC are both programmed to Yes the shock configuration can be programmed to RV to Case or Case to RV or RV to SVC or SVC to RV or RV to Case SVC or Case SVC to RV RV to Ca...

Страница 37: ...ying that the lead port is not filled with any obstacle 4 Insert the screwdriver into the pre inserted screw socket of the appropriate port in order to allow excess air to bleed out and to make the in...

Страница 38: ...n the can and the heart Suture the casing connector to the muscle using the hole provided for this purpose in order to avoid potential migration of the device into the pectoral muscle 14 12 TESTS AND...

Страница 39: ...PMT Smoothing Rate Response When the magnet is removed the sensor rate is forced to the basic rate arrhythmia detection algorithms and sequential therapies are reinitialized therapies start with the l...

Страница 40: ...ial sensitivity setting of 0 2 mV compliance to the Cenelec standard 45502 2 2 is met for a maximum test signal amplitude of 8 V for the frequency of 60 MHz 0 4 mV complies with the standard for the w...

Страница 41: ...yarrhythmia onset Searching for a long cycle P And R based Arrhythmia Detection PARAD Additional arrhythmia classification criterion to improve identification of atrial fibrillation and avoid inapprop...

Страница 42: ...te ventricular inhibition if crosstalk occurs Protection against noise Allows the distinction between ventricular noise and ventricular fibrillation If the device senses ventricular noise the ventricu...

Страница 43: ...clinic The data is transmitted from the implant and the SMARTVIEW monitor a small transmitter placed in the patient home Implant data are first transmitted to the SMARTVIEW monitor via RF Data are th...

Страница 44: ...ased on the degree of severity for the patient 16 8 4 User website On the website the physician is able to consult and schedule the remote follow ups of their patient configure additional ways of bein...

Страница 45: ...e reset the magnet rate is equal to 87 bpm it will be updated within the next 24 hours Refer to the online help for a description of displayed warning and the necessity to contact Sorin for an evaluat...

Страница 46: ...FF one 42 J shock every 2 weeks Once the Recommended Replacement Time RRT point has been reached the device operates normally except that the charge time increases Under normal conditions and without...

Страница 47: ...y shock 17 5 DEFIBRILLATOR IDENTIFICATION The defibrillator can be interrogated and programmed via telemetry using the programming head interfaced with the Sorin dedicated programmer Position the prog...

Страница 48: ...er of patients of patients Number of events Events per device year a Deaths 0 0 0 0 Explants 0 0 0 0 Serious pacemaker related events outside the use of SafeR AAI DDD 0 0 0 0 Non serious pacemaker rel...

Страница 49: ...mber rate responsive pacemaker with SafeR AAI DDD mode A variety of marketed atrial and ventricular pacing leads were used The pacemaker was programmed and interrogated via bi directional telemetry us...

Страница 50: ...of time that the patient spends in AV block SafeR AAI DDD cannot and should not provide any decrease in ventricular pacing while the patient is in AV block 1 Pioger G Jauvert G Nitzsch R Pozzan J Laur...

Страница 51: ...rface area of casing 76 cm2 Connector Atrium SonR Right ventricle DF 4 Left ventricle IS 1 19 1 MATERIALS USED Active surface area of casing 99 pure titanium Connectors Polyurethane and silicone elast...

Страница 52: ...gies maximum voltages and delivered energies at 37 C 50 ohm load for the minimum low mean and maximum programmed energy values is as follows Stored energy J 0 5 10 20 34 42 V1 Volt 75 341 483 631 702...

Страница 53: ...R ON 6 0 years Biventricular pacing in DDD mode 1 in atrium 100 in both ventricles 500 ohm 3 5 V 0 35 ms 60 bpm one 42 J shock per quarter sensors OFF 4 2 years Biventricular pacing in DDD mode 15 in...

Страница 54: ...5 95 100 110 115 125 135 140 150 155 165 170 180 190 195 205 210 220 225 235 250 19 ms Exercise AV delay ms 30 40 45 55 65 70 80 85 95 100 110 115 125 135 140 150 155 165 170 180 190 195 205 210 220 2...

Страница 55: ...3 ms 1 Values are measured using a positive and negative triangular signal of 2 13 ms 2 The correlation between the programmed amplitudes the stored amplitudes and the mid pulse delivered amplitudes u...

Страница 56: ...nsed Offset ms 0 10 15 25 30 40 45 55 65 70 80 85 95 100 110 115 125 1 ms A amplitude V 1 1 5 2 2 5 3 3 5 4 4 5 5 6 20 A pulse width ms 0 12 0 25 0 35 0 5 0 6 0 75 0 85 1 10 RV amplitude V 1 1 5 2 2 5...

Страница 57: ...VT VF detection zone Fast VT VF ON VF ON Slow VT rate lower limit bpm From 100 to 200 by steps of 5 190 VT rate lower limit bpm 130 135 140 145 150 155 160 165 170 175 180 185 190 195 200 210 220 230...

Страница 58: ...celeration 6 13 19 25 31 38 44 50 Long cycle persistence extension cycles From 0 to 16 by steps of 1 10 Long cycle gap ms 15 30 45 65 80 95 110 125 140 155 170 190 205 Atrial monitoring Yes No 21 3 VE...

Страница 59: ...m Values ATP program OFF Burst Burst Scan Ramp Ramp Scan Number of sequences 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cycles in first sequence 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cycles added per sequence...

Страница 60: ...Values ATP program OFF Burst Burst Scan Ramp Ramp Scan Number of sequences 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cycles in first sequence 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cycles added per sequence 0...

Страница 61: ...of Max Shock 42 J 1 2 3 4 21 4 REMOTE ALERTS AND WARNINGS The device routinely performs security self checks and technical measurements to ensure system integrity When system integrity is found to be...

Страница 62: ...0 Abnormal LV lead high limit Ohm 1500 1750 2000 2500 3000 Abnormal RV coil impedance ON OFF Abnormal SVC coil impedance ON OFF Abnormal Shock impedance 1 ON OFF 1 Normal impedance range 20 Ohm 200 Oh...

Страница 63: ...s Post ventricular pacing 220 ms 4 ms Post atrial pacing blanking 16 ms 3 ms Tachycardia criteria Values Window of PR association 63 ms 1 ms Therapies Values Waveform 1 Constant tilt 50 50 Stored ener...

Страница 64: ...erter defibrillator must be implanted prior to the use before date indicated on the packaging 4 The limited guarantee only applies to suspect devices returned to the manufacturer carefully packed and...

Страница 65: ...iod starting from the implantation date Sorin is committed to replacing free of charge the explanted device by a Sorin device with equivalent features or issuing a replacement credit equal to the purc...

Страница 66: ...15 5 776 164 5 776 165 5 818 703 5 836 980 5 868 793 5 891 170 5 891 184 5 899 931 5 931 856 5 935 153 5 954 660 5 978 708 6 181 968 6 230 058 6 236 111 6 251 703 6 256 206 6 307 261 6 337 996 6 397 1...

Страница 67: ...e oxide Temperature limitation High voltage Consult instruction for use This icon is used to call your attention to a particularly important point This icon alerts you to a hazard that may result in e...

Страница 68: ...CRM USA Inc Via Crescentino s n 14401 West 65th Way 13040 Saluggia VC Italy Arvada CO 80004 USA Tel 39 0161 487095 Tel 877 663 7674 Facsimile 39 0161 487524 www sorin com 2013 09 U150A CAUTION FEDERA...