– 5 –
KV-ES38N90
RM-958
2. DISPLAY OF STANDBY/TIMER LIGHT FLASH COUNT
Lamp ON 0.3 sec.
Lamp OFF 3 sec.
Lamp OFF 0.3 sec.
2 times
3 times
4 times
5 times
6 times
7 times
3. STOPPING THE STANDBY/TIMER FLASH
Turn off the power switch on the TV main unit or unplug the power cord from the outlet to stop the STANDBY/TIMER
lamp from flashing.
4. SELF-DIAGNOSTIC SCREEN DISPLAY
For errors with symptoms such as “power sometimes shuts off” or “screen sometimes goes out” that cannot be con-
firmed, it is possible to bring up past occurances of failure for confirmation on the screen:
[To Bring Up Screen Test]
In standby mode, press buttons on the remote commander sequentially in rapid succession as shown below:
[Screen display]
/
channel
[5]
/
Sound volume
[-]
/
Power ON
˘
Note that this differs from entering the service mode (mode volume
[+]
).
Self-Diagnosis screen display
STANDBY/SLEEP lamp
Diagnostic Item
Flash Count*
+B overcurrent
2 times
+B overvoltage
3 times
V deflection stop
4 times
White balance failure
5 times
High voltage protector
6 times
Audio Protection
7 times
* One flash count is not used for self-diagnostic.
002 : 000
003 : 000
004 : 000
005 : 001
Numeral "0" means that no fault has been detected.
Numeral "1" means a fault has been detected.
006 : 002
007 : 000
101 : 000
SELF DIAGNOSTIC
Numeral "2" means two faults have been detected.
Содержание TRINITRON KV-ES38N90
Страница 98: ... 122 KV ES38N90 RM 958 MEMO ...