SEL1018 (E 4/10-18 OSS) (E 10-18mm F4 OSS)
4-4
4. Turn the focus ring of the checking lens while observing the microscope and stop it at the position where the chart image is in focus.
Preparations of Using a B-L Control Box
Equipment and Tools
• Lens adjustment program (BLG-TestProgramV[][][].exe)
• B-L Control Box (Supplied with D-sub 15-pin cable)
• AC cable (Sold on the market)
• RS-232C Cable (D-sub 9-pin (Male with securing screws), D-sub 9-pin (Female with securing screws)) Straight type (Sold on the market)
• USB to serial cable (D-sub 9-pin (Male)) (Sold on the market)
• Personal Computer
Note:
Personal computer must have the condition that Windows Vista/7 runs normally and USB is connectable.
Setting Procedure of Equipment and Tools
1. Connect the B-L Control Box with the A-E mount conversion attachment and A-mount attachment, and connect it to the flange back tester.
2. Attach the extension jig (47.00 mm) to the objective lens (10x) to install it to the flange back tester.
3. Attach the checking lens to the flange back tester, and set the 1000 mm collimator.
1000 mm collimator
Checking lens
Extension Jig (47.00 mm)
A-E mount conversion
attachment
USB to Serial Cable
D-sub 9-pin: Female
RS-232C Cable
AC Cable
D-sub 9-pin:
Male
B-L Control Box
Personal computer
D-sub 15-pin
cable
Camera Table
200×200
FB Tester Stand
4. Install the BLG-TestProgram in the following procedure.
(1) Download the file “BLG-TestProgramV[][][].zip”, save and extract it.
(2) Save “BLG-TestProgramV[][][].exe” in an any folder.
Note:
The version of “BLG-TestProgram” may be updated without notice.
5. When starting and finishing the BLG-TestProgram, perform the following procedure.
• Starting the BLG-TestProgram
Starting Procedure
(1) Mount the checking lens to the lens mount block.
(2) Connect the D-sub 15-pin cable to the lens mount block, and turn on the B-L Control Box.
(3) Start “BLG-TestProgramV[][][].exe”.
(4) Click [Open] to connect communication with the B-L Control Box.
Note:
When the B-L Control Box or personal computer to be used is started for the first time, a password must be entered.
When the password input window opens, ask HQ about password. In this case, the ID shown in “COM / USER Characteristic Number” must be
informed
(“YdDGAAHTJ” in the example above).
Since this ID number is unique to the B-L Control Box and personal computer, if a password is used for another B-L Control Box or personal computer,
another password is necessary.
(5) Connection to the lens mount block is established and the [Power] button is made active.
• Finishing the BLG-TestProgram
Finishing
Procedure
(1) Click [Power] to turn off the power to the lens.
(2) Confirm that the two LEDs on the B-L Control Box are not lit.
LEDs for power supply check
B-L Control Box
(3) Click [Close] to disconnect communication between personal computer and B-L Control Box.
(4) Click [Finish] to exit the program.
6. Start the lens adjustment program “BLG-TestProgramV[][][].exe”.
7. When [Open] is clicked and communication between personal computer and B-L Control Box is connected, the button name changes to [Close]
and the [Power] button is made active.