![Sony LE-SP6800 Series Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/sony/le-sp6800-series/le-sp6800-series_safety-manual_410998010.webp)
10
Specifications
Specifications
General
Optics System
Measurement system
Configuration
2-laser type (488/638)
2-laser type (405/488)
3-laser type (405/488/638)
Model
LE-SP6800ZB
LE-SP6800ZC
LE-SP6800ZE
Dimensions
Main unit: 60.0 cm × 63.5 cm × 71.3 cm (23
5
/
8
in. × 25 in. × 28
1
/
8
in.) (w/d/h)
Fluidics cart: 78.6 cm × 52.1 cm × 58.0 cm (31 in. × 20
5
/
8
in. × 22
7
/
8
in.) (w/d/h)
Mass
Main unit: Approx. 99 kg (218 lb. 4 oz.) (dry weight),
Fluidics cart: Approx. 32 kg (70 lb. 8 oz.) (dry weight)
Compressed air supply
pressure
350 kPa to 450 kPa (51 psi to 65 psi)
Power supply
100 V to 240 V AC, 50/60 Hz
Power consumption
220 W (maximum)
Operating temperature
16 °C to 29 °C (61 °F to 84 °F), room temperature variation: < 5 °C (9 °F)
Operating humidity
20% to 80% (condensation free)
Configuration
2-laser type (488/638)
2-laser type (405/488)
3-laser type (405/488/638)
Model
LE-SP6800ZB
LE-SP6800ZC
LE-SP6800ZE
Laser specifications
Wavelength
488 nm, 638 nm
semiconductor lasers
405 nm, 488 nm
semiconductor lasers
405 nm, 488 nm, 638 nm
semiconductor lasers
Maximum output
(irradiation power at flow
cell chip)
488 nm: 40 mW
638 nm: 60 mW
405 nm: 60 mW
488 nm: 40 mW
405 nm: 60 mW
488 nm: 40 mW
638 nm: 60 mW
Irradiation method
2-spot dual-axis irradiation
Detection system
Scatter measurement
Forward scatter (FSC), Side scatter (SSC)
Spectroscopic method
Prism array spectroscopy
Fluorescence
measurement
32-channel PMT
(500 nm to 800 nm detection
wavelength)
32-channel PMT (500 nm to 800 nm detection wavelength)
PMT × 2 (420 nm to 440 nm, 450 nm to 470 nm)
Cell passage position
measurement
4-segment photodiode (4 optical sensors)
Configuration
2-laser type (488/638)
2-laser type (405/488)
3-laser type (405/488/638)
Model
LE-SP6800ZB
LE-SP6800ZC
LE-SP6800ZE
Signal resolution
Height: 20 bits, Area: 32 bits (equivalent to 4.29 million channels),
Sampling frequency: 50 MHz
Measurement parameters
64 color parameters, FSC,
SSC, cell passage position
coordinates (X,Y)
66 color parameters, FSC, SSC,
cell passage position coordinates (X, Y)
Pulse processing
Height, Area, Width (all measured)
Optical axis adjustment
method
Automatic adjustment using CoreFinder™ technology
Event rate
10,000 eps (nominal), 20,000 eps (maximum)