3
This unit can not be used in the patient
environment.
3. Do not touch the patient simultaneously
while you are contacting with this
equipment.
4. For this particular equipment, all
accessory equipment connected as noted
above, must be connected to mains via an
additional isolation transformer
conforming with the construction
requirements of IEC60601-1 and
providing at least Basic Insulation.
5. The leakage current could increase when
connected to other equipment.
6. This equipment generates, uses, and can
radiate radio frequency energy. If it is not
installed and used in accordance with the
instruction manual, it may cause
interference to other equipment. If this
unit causes interference (which can be
determined by unplugging the power
cord from the unit), try these measures:
Relocate the unit with respect to the
susceptible equipment. Plug this unit and
the susceptible equipment into different
branch circuit.
Consult your dealer. (According to standard
EN60601-1-2 and CISPR11, Class B,
Group 1)
Caution
When you dispose of the unit or
accessories, you must obey the laws in the
relative area or country and the
regulations in the relative hospital.
R 1.5 m
* Patient Environment.