Command Descriptions
3026 Programmer Manual
2Ć71
TRIGger:LEVel (?)
If the trigger is to be generated by a signal above a particular amplitude level in
the time domain, the
TRIGger:LEVel
command is used to set the level. The
TRIGger:LEVel?
query returns the current trigger level setting.
TRIGGER
TRIGger:SOURce
TRIGger:LEVel <numeric value>
TRIGger:LEVel?
<numeric value>::=<NR2>
where
<NR2>
is a trigger level value ranging from 0.000 V to 8.000 V.
<NR2>
:TRIGGER:LEVEL 0.01
sets the level to 0.01 V.
TRIGger:MODE (?)
The
TRIGger:MODE
command sets the trigger mode. The
TRIGger:MODE?
query
returns the current trigger mode setting.
TRIGGER
None
TRIGger:MODE {NORMal|AUTO}
TRIGger:SLOPe?
NORMal
selects the Normal mode.
AUTO
selects the Auto mode.
:TRIGGER:MODE AUTO
sets the trigger mode to Auto.
Group
Related Commands
Syntax
Arguments
Responses
Examples
Group
Related Commands
Syntax
Arguments
Examples
Содержание 3026
Страница 6: ...Table of Contents 3026 Programmer Manual ...
Страница 8: ...Preface 3026 Programmer Manual ...
Страница 9: ... ...
Страница 10: ......
Страница 15: ... ...
Страница 16: ......
Страница 24: ...Syntax 3026 Programmer Manual ...
Страница 90: ...Command Descriptions 3026 Programmer Manual ...
Страница 91: ... ...
Страница 92: ......
Страница 103: ...Error Messages and Codes 3026 Programmer Manual ...
Страница 105: ...Error Messages and Codes 3026 Programmer Manual These error codes are returned in response to an unanswered query ...
Страница 106: ...Error Messages and Codes 3026 Programmer Manual ...
Страница 107: ... ...
Страница 108: ......
Страница 112: ...Appendix B Reserved Words 3026 Programmer Manual ...
Страница 116: ...Appendix C Interface Specification 3026 Programmer Manual ...
Страница 119: ... ...
Страница 120: ......
Страница 125: ......
Страница 126: ......