Main System Disassembly for Repair and/or Replacement
Required Parts
Parts for the Main System Repair could include any of the following:
• P08939 Service Assembly Main PCBA, M-Turbo
• P08850 Service Assembly Power Supply, M-Turbo
• P05470 Service Assembly TGC, MicroMaxx (compatible with MicroMaxx and M-Turbo)
• P05473 Service Assembly Speaker, M-Turbo
• Nest Frame Assembly, M-Turbo (order these parts individually as necessary)
• P00364 Connector, Interposer (Qty 8)
• P00924 Screw, Shoulder, Thrust Plate (Qty 4)
• P00353 Wear Plate
• P00646 Spring, Thrust Plate (Qty 4)
• P07750 Nest Frame
• P03834 Shield, Perimeter, Long (Qty 2)
• P03833 Shield, Perimeter, Short (Qty 2)
• P08200 M2.5-.45x10 Socket Head Cap Screw (Qty 4)
Required Tools
• #1 Phillips screwdriver
• Torque screwdriver, 2.0–10.0 inch pounds (0.23–1.1 newton meter)
• 2 mm allen key
• Scissors
• Q-Tips
• An anti-static mat
• A wrist grounding strap
System Disassembly
Caution:
Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first.
Initial ESD symptoms may be slightly degraded performance or image quality.
System
Disassembly
1
Remove the battery.
2
Remove the control panel from the system following the removal procedures in
.
3
Remove the 4 remaining screws from the bottom of the system.
4
Remove the bottom enclosure. This exposes all of the replaceable parts for the main system
per
.
Содержание M-Turbo
Страница 1: ...M Turbo Ultrasound System Service Manual TM...
Страница 6: ...2 Chapter 1 Introduction...
Страница 20: ...16 Chapter 3 Troubleshooting...
Страница 36: ...32 Chapter 4 Replacement Procedures...
Страница 44: ...40 Chapter 5 Performance Testing...
Страница 48: ...44 Appendix A Replacement Parts List Figure A 1 Power Supply P08850 Figure A 2 Speaker Assembly P03872...
Страница 52: ...48 Appendix B Service Event Report Service Event Report Form...
Страница 56: ...52 Index...
Страница 57: ......
Страница 58: ...P08144 01 P08144 01...